Specification
Unit
Value
Weight
kg
59
Noise level
dB(A)
No increase in the noise level of the constant
temperature equipment
Power supply via LiBus cable
VDC
24
Power consumption
kW
0.02
Distance between device and environment
‑
Front
‑
Back
‑
Right
‑
Left
mm
mm
mm
mm
200
200
200
200
The maximum flow rate through the flow controller depends on
n
the maximum flow rate of the constant temperature equipment,
n
the temperature of the heat transfer liquid,
n
the pressure drop in the application and the hoses.
Table 11: Maximum flow rate depending on the temperature, reference devices measured with Kryo 30. Lower values may
result depending on the hydraulic resistance in the system.
Temperature tb [°C] of the
Heat transfer liquid
IN 750 XT
IN 2050 PW
Unit for flow rate
140
---
88
l/min
120
---
87
l/min
90
54
84
l/min
60
52
81
l/min
40
51
79
l/min
20
48
77
l/min
0
44
71
l/min
-10
40
66
l/min
-20
34
59
l/min
-30
25
45
l/min
-40
12
25
l/min
The characteristics serve as a guide
The control accuracy depends on the outflow pressure, which in
turn results from the resistance of the connected application.
As a general rule:
The lower the flow rate, the higher the control accuracy.
The lower the temperature, the higher the control accu-
racy.
Characteristics for IN 2050 PW and
IN 750 XT
V2
Flow controller MID 80
32 / 38
Содержание MID 80
Страница 34: ...12 Declaration of Incorporation V2 Flow controller MID 80 34 38 ...
Страница 38: ...Warranty 30 V2 Flow controller MID 80 38 38 ...
Страница 39: ......