![Lauda MC 250 Скачать руководство пользователя страница 52](http://html2.mh-extra.com/html/lauda/mc-250/mc-250_operation-instructions-manual_3834611052.webp)
11
Technical Data
11.1
General data
The device sound pressure level is below 70 dB. According to EC
Directive 2006/42/EC. the sound pressure level of the devices is
therefore not specified further.
Specification
Value Unit
Placement
Indoor areas
Placement height above sea level
Up to 2,000 m
Humidity
Maximum relative humidity 80% at 31 °C and up to
40 °C decreasing linearly to 50%
Ambient temperature range
5 – 40 °C
IP type of protection
IP 32
Degree of soiling
2
Clearance from surroundings (front and rear sides)
40 cm
Overvoltage
Overvoltage category II and transient overvoltages
according to category II
Protection class for electrical operating equipment
DIN EN 61 140 (VDE 0140-1)
1
Classification according to DIN 12 876-1 (class
designation / identification)
I/NFL
Display
7-segment, LED
Display resolution
0.1 °C
Adjustment resolution
0.1 °C
Temperature stability
±
0.5 K
Storage temperature range
5 – 40 °C
Transport temperature range
-20 – 60 °C
Working temperature range
Dimensions (W x D x H)
Weight
°C
mm x mm x mm
kg
MC 250
-10 – 40
200 x 350 x 465
28
MC 350
-10 – 40
240 x 400 x 500
36
MC 600
-10 – 40
350 x 480 x 595
52
MC 1200
-10 – 40
450 x 550 x 650
64