Integral XT
YAWE0028 / 24/06/2016
Starting up
57
7.9
Important settings
7.9.1 Temperature setpoint setting
The setpoint is the temperature which the thermostat should reach and maintain constant.
Master
(main level)
SEt
−
Press key until
SEt
(Setpoint) appears.
−
Press, display flashes.
or
−
Enter the setpoint with the two keys ((
þ
7.4.1) General key func-
tions and pilot lamps).
Wait 4 seconds or
−
Display flashes 4 s
à
new value is automatically accepted,
or
value is accepted immediately with Enter key.
+
−
For safety reasons the setpoint can only be set up to 2 °C above
upper limit of the operating temperature range for the relevant de-
vice type.
−
In the following cases the manual setpoint entry is blocked:
Setpoint is taken from the analog module, from the programmer in
the Command remote control or via the serial interface.
−
When the setpoint temperature is to be lowered, it may take up to
one minute before the blue LED
lights.
Содержание Integral XT 150