5.5 Fill the device with heat transfer liquid
LAUDA is not liable for damages resulting from the use of unsuitable heat
transfer liquids.
Do not mix different heat transfer liquids together!
Use a funnel if required to fill the device with heat transfer liquid.
The filling of a device with a pressure pump (mandatory in IN 2560 XTW) is
described in
“Filling with a pressure pump” on page 74.
The filling of devices with pressure overlay is described in
DANGER!
Use of unsuitable heat transfer liquid
Fire
Select a heat transfer liquid with a temperature range suit-
able for the application.
WARNING!
Overheating of the heat transfer liquid
Fire
The heat transfer liquid used in the device must be set in
the menu of the device.
WARNING!
Risk of heat transfer liquid overheating due to incorrect entry of
overtemperature switch-off point Tmax
Fire
Set the overtemperature switch-off point to 5 K above the
upper limit of the temperature range for your application.
The overtemperature switch-off point must be below the
flash point of the heat transfer liquid.
WARNING!
Splashing heat transfer liquid
Eye damage
Always wear suitable safety glasses when working on the
device.
V6
Integral Process Thermostats and High-Temperature Thermostats
71 / 198
Содержание IN 130 T
Страница 89: ...6 3 Menu structure Fig 45 Menu Part 1 V6 Integral Process Thermostats and High Temperature Thermostats 89 198...
Страница 182: ...15 9 Declaration of Conformity and certificates V6 Integral Process Thermostats and High Temperature Thermostats 182 198...
Страница 183: ...Fig 99 Declaration of Conformity V6 Integral Process Thermostats and High Temperature Thermostats 183 198...
Страница 184: ...Fig 100 Declaration of Conformity V6 Integral Process Thermostats and High Temperature Thermostats 184 198...
Страница 186: ...V6 Integral Process Thermostats and High Temperature Thermostats 186 198...
Страница 187: ...V6 Integral Process Thermostats and High Temperature Thermostats 187 198...
Страница 199: ......