Emptying
−
Switch off the thermostat, pull out the mains plug!
a) Immersion thermostat
−
Unscrew the thermostat-
−
Drain the bath
b) Bath/Circulation thermostat
−
Take off the control unit with the bath bridge
−
Drain the bath.
☞
−
The units are designed for operation with non-flammable and flammable liquids to EN
61010-2-010! Flammable liquids can be operated up to no more than 25°C below the
firepoint (
➾
Section 5.3.).
−
Observe the appropriate regulation when disposing used thermostating liquid.
−
When connecting an external consumer take care that the level of the bath liquid does
not drop too much
!
fill in bath liquid if necessary.
Do not drain the thermostating liquid when it is hot or very cold (below 0°C)!