Starting up
15.05.01
YAEE0013
31
−
with key back to "Selection of the interface".
6.4.3.5 User
calibration
☞
−
Remove the external consumers and switch the setting knob of the pump to right side.
(
➾
Section 5.4).
−
A reference thermometer with necessary accuracy is required. Otherwise the factory
calibration should not be altered. The reference thermometer has to be inserted far
enough and long enough into the bath.
−
It is not allowed to calibrate to more than ± 3 °C. Multiple calibration to more than ± 3
°C cause internal faults (after 2 min ”E1006” or ”e16”).
−
The factory calibration will be lost through overwriting
!
!
!
!
please work
carefully!!!
and
2x
−
Directly from level 0 (setpoint selection), press the key
combination on the left, until the PARA-function is reached,
then
and
4x
−
switch to level 2 and move forward with key on the left.
>2,5 s
1. The display shows CAL and the current bath temperature.
To carry out a calibration, press the key longer than 2.5 s.
2. The actual value appears.
or
3. Input the value indicated on the reference thermometer with
one to the two keys.
4. The additive calibration must be entered with the key shown
on the left.