21
4. MOUNTING PRECAUTIONS
4-1. When Mounting
•
The scale and the detector head are adjusted to each
other.
If more than one scale unit is to be used, only combine
them with the detector heads provided in the same
packing. If you do not know which scale and detector head
to use together, refer to the No. on the packing carton.
•
Consider the mounting position (Abbe error) and the
environmental conditions (temperature, humidity,
oscillation and dust) thoroughly.
•
Do not lead the head connector cable through the same
duct with the power cable.
•
Set the maximum travel of the machine’s moving part
shorter than that of the scale.
Scale’s max. trave : effective measuring length
+
10 mm
/0.39" (5 mm/0.20" per each side)
(If the scale’s max. travel is exceeded, the signals of the
scale unit will not be output and an error will occur.)
•
For the scale unit, scale signal adjustment is necessary
after mounting.
When mounting the scale to the machine, make sure that
there is enough space for the adjustment.
(see page 23 to 32)
•
To protect the scale against external light, dust, oil, etc.,
only use it with the cover.
•
Be sure to set the display unit or detector to the
compensation value of the scale before operation.
4-2. Mounting Requirements
Measuring Instruments
•
Lever type dial gauge (scale pitch: 2/1000 mm/0.00008"
or 1/1000 mm/0.00004") ............ 1
•
Dial gauge stand ............................................................. 1
•
Oscilloscope ................................................................... 1
Capable of 2-quadrant display
Input sensitivity
: DC 0.1 V/DIV
X-Y frequency band : 1 MHz or more
Tools
•
Hexagonal wrench
4 mm/0.157", 3 mm/0.118", 2 mm/0.079" ............... 1 each
•
Philips-head screwdriver No.1 ........................................ 1
Miscellaneous
•
Accessories ............................................................... 1 set
•
Gauze ......................................................... Small amount
•
Alcohol ........................................................ Small amount
Содержание BS75A-R
Страница 4: ...2 1 1 100 2 780nm 3mW 5 ON ON 50mm 30...
Страница 5: ...3 No MADE IN JAPAN No MADE IN JAPAN 3 M4 8 M5 40...
Страница 6: ...4 4 4 1 No 10mm 5mm 6 14 4 2 1 2 1000 mm 1 1000 mm 1 1 2 X Y DC 0 1V DIV X Y 1 MHz 4 3 2 1 1 1 1...
Страница 13: ...11 No MADE IN JAPAN u 1 4 N m BS75A 260R RS 310R RS BS75A 360R RS 410R RS i 5 3 q w M5 40 5 B A A B B A A B...
Страница 15: ...13 No MADE IN JAPAN 5 5 M4 8 3 5 6 q w 50mm 14 e r 0 2mm S M4 8...
Страница 18: ......
Страница 36: ......
Страница 54: ......