2-For connection to your amplifying electronics use the existing pigtail exiting from one side of the 

head. Signals  generated  by sensors are  at  very low voltage:  please take care to  keep the distance 
between sensor and electronics as short as possible.  

 

2.3 - Installation of Air Convection Cooled Series 

Mechanical mounting on site can be done by the 2 M3x 5mm deep threaded holes located on a 
lateral side of the head.. Always  take care to have the beam centered on the sensor area. 
Housings are optimized for the max power specified for the 
sensor, nevertheless it can be necessary to enhance dissipation. In 
situations like poorly vented cabinets or when the head faces air 
streams from fans, it is advisable to fix the head to suitable 
additional heat sink: 

1)

 

A custom heat sink may be installed either onto the 
existing rear flange or , more efficiently, directly against 
the main body of housing after the 4x M3 screws and  the 
rear flange have been removed. 

2)

 

Always maintain a suitable layer of heat transfer 
compound between the rear flange and the head  body to increase thermal conductivity. 

2.4 - Installation of Water Cooled Series 

Mechanical mounting on site can be done by the 2 M3x 5mm deep threaded holes located on a 
lateral side of the head.  
When mounting always  take care to have the beam 
centered on the sensor area. 
1-Secure the resulting detector assembly to a suitable 
holder (eg. a right angle bracket or plate) keeping the 
center of the sensor coaxial to the laser beam. 
Mounting can be done by 2 x M3 screws 
2-Insert 6mm O.D.- 4mm I.D. Polyurethane hoses in 
the Fittings; Water flux capability of each head has 
been optimized for the max power sustainable by the 
sensor: check for proper water flux on the OEM sensors specs (refer to Section 5 for specifications 
on optimal water flux). 

 
 
 
 

 
  
 

 
 
 
 

 
 
 
 
 
 

Содержание CSA-20-D12

Страница 1: ...e Via Burona 51 tel 02 27 40 02 36 fax 02 25 02 91 61 e mail sales laserpoint it www laserpoint eu 1 06 T Th he er rm ma al l O OE EM M L La as se er r S Se en ns so or rs s M Ma an nu ua al l Instruction Manual Edition 2 July 2017 ...

Страница 2: ...LE OF CONTENTS TABLE OF CONTENTS 2 1 GENERAL 3 2 INSTALLING THE CS OEM SENSOR HEAD SERIES 4 2 1 General 4 2 2 Output Signal Condictioning 4 2 3 Installation of Air Convection Cooled Series 5 2 4 Installation of Water Cooled Series 5 3 STANDARD OEM SENSOR HEADS 6 Standard Thermal Heads 6 4 SAFETY 9 5 WARRANTY 9 6 LIMITED LIABILITY 10 ...

Страница 3: ...pile detector units based on a sensor integrated in a suitable mechanical housing that according to the power sustainable by the sensor are either convection air cooled CSA or water cooled CSW Unpacking and inspection Each OEM Detector Head has been carefully inspected and tested prior to shipment Unpack and assure that no damage occurred during transport and handling Any damage must be immediatel...

Страница 4: ...or the absorbing coating If the absorber is damaged measurements may be affected and a substitution of the sensor may be necessary refer to Section 4 for individual threshold of each sensor 2 2 Output Signal Condictioning External connection options include Laser Point s IIS Intelligent Identification System on DB15 connector BNC 4 wire pigtail 2 2 1 The DB15 Connector Heads with DB 15 male connec...

Страница 5: ...table additional heat sink 1 A custom heat sink may be installed either onto the existing rear flange or more efficiently directly against the main body of housing after the 4x M3 screws and the rear flange have been removed 2 Always maintain a suitable layer of heat transfer compound between the rear flange and the head body to increase thermal conductivity 2 4 Installation of Water Cooled Series...

Страница 6: ...nto 100K Load 1 95 mV W 20 batch dependent NEP Noise Equivalent Power 2 5 mW Response Time 0 90 1 sec b Linearity 1 c Cooling Conduction by heat sink Notes a Tested 1 06 µm b Accelerated by LaserPoint electronics c Detector centrally irradiated 50 of useful surface CSA 20 D20 TYPICAL LASERS Diode Lasers Diode arrays CO2 Nd Yag Ion Lasers Max Continuous Power 20 W Max Energy 20 J long pulses Outer ...

Страница 7: ... 0 90 1 sec b Linearity 1 c Cooling Water 0 5lit min 22 C Notes a Tested 1 06 µm b Accelerated by LaserPoint electronics c Detector centrally irradiated 50 of useful surface CSW 50 D25 TYPICAL LASERS Diode Lasers Diode arrays CO2 Nd Yag Ion Lasers Max Continuous Power 50 W Max Energy 50 J long pulses Outer Dimensions 60x60x20mm Sensitive Diameter 25mm Available Absorber Types HCB Hard Coating Broa...

Страница 8: ...ec b Linearity 1 c Cooling Water 1 5lit min 22 C Notes a Tested at 10 6µm b Accelerated by LaserPoint electronics c Detector centrally irradiated 50 of useful surface CSW 200 D25 TYPICAL LASERS Diode Lasers Diode arrays CO2 Nd Yag Ion Lasers Max Continuous Power 200 W Max Energy 200 J long pulses Outer Dimensions 60x60x23mm Sensitive Diameter 25mm Available Absorber Types HCB Hard Coating Broadban...

Страница 9: ... warranty from date of shipment Warranty applies against material and or workmanship defects whether used under normal operating conditions the warranty does not cover damages related to accident or misuse During the warranty period LaserPoint srl will repair or at its sole option replace any OEM sensor head or its components that prove to be defective provided the parts are returned to LaserPoint...

Страница 10: ...ermal OEM Laser Sensors are products compliant to RoHS European Directive 7 European Union WEEE Directive WEEE 2002 96 EC The European Waste Electrical and Electronic Equipment WEEE Directive 2002 96 EC is represented by a crossed out garbage container label see below The purpose of this directive is to avoid the disposal of WEEE as unsorted municipal waste and to facilitate its separate collectio...

Отзывы: