18
EN
Checking the horizontal line
Position the device about 5 m from a wall and
switch on the cross laser. Mark point B on the
wall. Turn the laser cross approx. 2.5 m to the
right and mark point C. Check whether the
horizontal line from point C is level with point B
to within ± 1 mm. Repeat the process by turning the laser to the left.
B
C
2,5 m
< 1 mm = OK
Checking the vertical line
Position the device about 5 m from a wall. Fix a plumb bob with a line of 2.5 m
length on the wall, making sure that the bob can swing freely. Switch on the
device and align the vertical laser to the plumb line. The precision is within the
specified tolerance if the deviation between the laser line and the plumb line
is not greater than ± 1 mm.
Regularly check the calibration before use, after transport and after
extended periods of storage.
!
Calibration
The meter needs to be calibrated on a regular basis to ensure it produces
accurate measurement results. We recommend carrying out calibration
once a year.
Information on maintenance and care
Clean all components with a damp cloth and do not use cleaning agents,
scouring agents and solvents. Remove the battery pack before storing for
longer periods. Store the device in a clean and dry place.