27
CoatingTest-Master
11.1
11.2
Checking calibration
1
Zero calibration; see item 10
2
/K@BDSGDQDPTHQDCQDEDQDMBDÖKLRNMSGDTMBN@SDC
material (original material or reference material)
3
Measure the reference thickness
Value within tolerance
(see technical data):
Check complete;
no one-point or two-point calibration required
Value outside tolerance
(see technical data):
One-point and two-point calibration required; see item 11.2
Perform one-point and two-point calibration
1
Zero calibration; see item 10
One-point calibration
(cal 1)
2
Turn the device off/on
3
Set the measuring mode:
Menu > Options > Measure mode > Single mode
Exit the menu: press key 1 twice
4
Set the working mode:
Menu > Options > Working mode > Direct / Group 1 - 4
"@KHAQ@SHNMB@MADCDÖMDCHMCHUHCT@KKXENQD@BGLNCD
Exit the menu: press key 1 twice
5
Delete the previous calibration data in the applicable working mode:
see item 7
6
Set the calibration mode:
Menu > Calibration > Enable > ESC
(„cal“ and „zero“ appear on the display)
7
Place the required reference film(s) on the uncoated
material (original material or reference material)
8
Measure the reference thickness
9
Set the reference film thickness using the „
“/„
“ keys
10
!QHD×XOQDRRSGD9$1.JDX
(„cal 1“ and „zero y“ appear on the display)
EN