(Recommend performing calibration check once a week at minimum) 

 

With the Quad 1000HDD mounted in place on the mounting trivet plate.

 

 
1.

 

Place the target / template precisely centered on the laser spot at the lowest point possible below 
the Quad 1000HDD. (At least 100 feet if possible) Secure the target so it will not move.

 

 
2.

 

Rotate the Quad 1000HDD on its rotational mount 180°. 

 

 
3.

 

Go back to the target and measure any change of the laser spot relative to the bulls-eye target.

 

 

 Given the dynamics and movement of structures due to temperature, wind, loads etc., the target 
relation to the Quad 1000HDD laser may be moving and not providing a constant. This should definitely 
be taken into consideration so as a result, depending on conditions, you could be getting readings of 

 

1/16 or 

 1/8 at the target per 100 vertical feet. If you are not comfortable or confident with the field 

calibration check, have it checked by your Dealer Service Department. 

 

 

 
 
 
MAINTENANCE

 

The Quad 1000HDD requires no maintenance other than periodically cleaning the laser output lens. Use 
a soft cloth with Windex or alcohol. 

*NOTE: Laser must be turned OFF when cleaning the lens! 

 
 

CERTIFICATION 

USA EN/IEC 60824-1 2007 Class 3R 
 
 

LASER SAFETY 
Laser Light-Avoid Direct Eye Exposure 

The maximum power output of this laser is less than 5mw. A laser safety kit is supplied with the Quad 
1000HDD laser. The kit contains operator qualification cards and a sign that should be posted near the 
laser whenever it is in use. Observe the following rules: 

 

Never look directly into a laser beam or point the beam into the eyes of others. Set the laser at a 
height that prevents the beam from shining directly into people’s eyes. 

 

DO NOT remove any warning signs from the laser. 

 

Only properly trained people are to use this product. 

 

Service to be performed only by factory trained personnel; unit contains no user serviceable 
parts. 

 

 

 

 

JOBSITE CALIBRATION CHECK 

Содержание 4003-0000

Страница 1: ...1041 Rev 2 28 19 1810 SE First Street Suite H Redmond OR 97756 Phone 541 548 0882 Fax 541 548 0892 Email laserline laserline net Website www laserlinemfg com VERTICAL CONTROL LASER OWNERS MANUAL Rev...

Страница 2: ...4001 0010 Ultra bright bulls eye for the most visible and precise readings at any distance Targets feature laser pass through allowing multiple readings at one time special hard coating that protects...

Страница 3: ...of 2 000 feet 610 meters Rugged Heavy duty inch thick aluminum housing with hard coat anodized finish Dust and waterproof Purged and dry nitrogen filled with IP67 rating OVERVIEW 5 8 Female thread for...

Страница 4: ...e top of the Quad 1000HDD Level the bulls eye bubble using the three leveling screws and traverse the Quad 1000HDD across the platform using a total station to position the unit to the plumb point Onc...

Страница 5: ...latform with oversized hole Flat Level Rigid Laser Trivet Mount Position laser over or under plumb point Make sure bulls eye bubble remains centered using the three adjustment Knobs Once in place lock...

Страница 6: ...he target center that allows the center of the laser spot to pass through so that more than one target can be used at the same time E There are two mounting holes in each target to allow for custom mo...

Страница 7: ...fident with the field calibration check have it checked by your Dealer Service Department MAINTENANCE The Quad 1000HDD requires no maintenance other than periodically cleaning the laser output lens Us...

Страница 8: ...Identification and Certification Non Interlocked Warning Label Top Explanatory and Warning Label Hazard Symbol Aperture Label Bottom LABELS Required for this product...

Страница 9: ...ss 3R International IEC 60825 1 2014 Class 3R USA OSHA compliant Meets all Federal and International standards Power source 12 VDC or 110 220 VAC with optional 1140 power converter Operating temperatu...

Страница 10: ...always check for accuracy per the enclosed instructions in this manual Warning Under no circumstances attempt to open or disassemble the Laser Housing Doing so may cause exposure to potentially hazar...

Страница 11: ...NOTES...

Отзывы: