Technical specifications for COUNT

®  

BLUE module

Parameter

Min.

Typ.

Max.

Unit

Spectral range

350

1000

nm

Dark count rate 
COUNT

®

-10B 

COUNT

®

-20B 

COUNT

®

-50B 

COUNT

®

-100B 

COUNT

®

-250B

 

10 
20 
50 

100 
250

 

Counts/s 
Counts/s 
Counts/s 
Counts/s 
Counts/s

Photon detection efficiency Pd

1

 at: 

405 nm 
520 nm 
670 nm

 

50 
60 
50

 

55 
70 
55

 



%

Pd variation with FC-connection

5

%

Pd variation at constant temperature

1

%

Dark count rate variation at constant temperature

5

%

Active area diameter (nominal) 

2

100

µm

Timing resolution 

3

1000

ps

Afterpulsing probability 

4

0.2

%

Dead time

42

45

48

ns

Gating input voltage 
Gating on (=disable module) 
Gating off (=enable module)

 

TTL low (<0.5) 

TTL high (>2.4)

 


V

Gating input response time 
Gating on (=disable module) 
Gating off (=enable module)

 

15 
60

 

20 
65

 

ns 
ns

TTL output pulse length

15

17

ns

TTL output pulse amplitude (into 50 Ohm)

3

V

Delay between photon impact and TTL pulse

30

ns

Supply voltage

11.5

12.0

12.5

V

Supply current (switch on)

0.8

A

Supply current (operation at 1 Mcps)

0.2

A

1

 Specifications valid for modules without FC connector. 

2

 The active area of the integrated Si-APD is larger than 100 µm. The FC-version is optimised for optical fibers with core diameter <105 µm. 

  The pre-aligned GRIN lens focuses the light onto a spot of < 70 µm diameter in the centre of the detector. The NA of the fiber used with the 
  module is recommended to be ≤ 0.15. For fibers with a NA ≤ 0.29 and a core diamenter of ≤ 105 μm special FC-versions are available 
  for wavelengths ≥ 500 nm. 

3

 Timing resolution depends on count rate and wavelength.  

4

 Defined for a time interval from 0 to 500 ns.

Single Photon Counting Module 
COUNT

®

-Series User Manual

Detectors

Germany and Other Countries

Laser Components Germany GmbH 
Tel:  +49  8142  2864 – 0 
Fax:  +49  8142  2864 – 11 
[email protected]

www.lasercomponents.com

Содержание COUNT Series

Страница 1: ...4 Optimizing performance 4 Performance 5 Spectral response 5 Timing resolution 5 Correction factor 6 Technical Data 7 Technical specifications for COUNT module 7 Technical sSpecifications for COUNT BLUE module 8 Technical specifications for COUNT NIR module 9 Absolute maximum ratings 10 Interface information 10 Individual module test data 10 ESD information 10 Safety 10 RoHS compliance 10 Warranty...

Страница 2: ...utput The optional FC connector provides a convenient method for connecting the module to the sample using a multimode optical fiber Mounting the Count modules The COUNT can be either mounted from the front side or by using the dedicated mounting holes on the base plate Front side Single Photon Counting Module COUNT Series User Manual The front side of the COUNT provides two mounting options 1 a C...

Страница 3: ...r mounting the module onto a suitable heat sink e g an optical table Gating function TTL output The COUNT module features a gating input provided as standard This can be used for viewing a signal that occurs only in a small timeframe window or in applications with very weak signals and higher background light The output of the module is disabled when a TTL low level is applied to the gate input Ap...

Страница 4: ...which the sensor will be cooled down to its operating temperature NOTE the COUNT will not generate any output signal until the operating temperature has been reached Optimizing performance The COUNT shows best performance if the light is focussed to a small spot 50 µm diameter in the center of the sensor area Please also see Fig 2 of the COUNT datasheet Mounting the COUNT on a suitable x y z trans...

Страница 5: ... timing resolution the light should be focused to a small spot 50 µm in the center of the sensor Off center focusing or overfilling the sensor area might lead to an increased FWHM of the photon timing resolution 3 Count rate the FWHM of the photon timing resolution increases at high count rates Especially at count rates 1 million counts second the FWHM may be double the value at low count rates Th...

Страница 6: ...be used to correct the non linearity especially at high light levels Rmeasured Rphoton 1 Rmeasured TD The following two figures demonstrate the effect of the dead time on the measured count rate and its correction factor Correction Factor R photon Rmeasured 1 1 Rmeasured TD Fig 2 Measured count rate vs actual photon rate for a dead time of 43 ns Fig 3 Correction factor vs photon rate for a dead ti...

Страница 7: ...65 ns ns TTL output pulse length 15 17 ns TTL output pulse amplitude into 50 Ohm 3 V Delay between photon impact and TTL pulse 30 ns Supply voltage 11 5 12 0 12 5 V Supply current switch on 0 8 A Supply current operation at 1 Mcps 0 2 A 1 Specifications valid for modules without FC connector 2 The active area of the integrated Si APD is larger than 100 µm The FC version is optimised for optical fi...

Страница 8: ... output pulse length 15 17 ns TTL output pulse amplitude into 50 Ohm 3 V Delay between photon impact and TTL pulse 30 ns Supply voltage 11 5 12 0 12 5 V Supply current switch on 0 8 A Supply current operation at 1 Mcps 0 2 A 1 Specifications valid for modules without FC connector 2 The active area of the integrated Si APD is larger than 100 µm The FC version is optimised for optical fibers with co...

Страница 9: ...odule Gating off enable module 15 60 20 65 ns ns TTL output pulse length 15 17 ns TTL output pulse amplitude into 50 Ohm 3 V Delay between photon impact and TTL pulse 30 ns Supply voltage 11 5 12 0 12 5 V Supply current switch on 0 8 A Supply current operation at 1 Mcps 0 2 A 1 Specifications valid for modules without FC connector 2 The active area of the integrated Si APD is larger than 100 µm Th...

Страница 10: ...robability ESD Information Modules should only be handled at an ESD safe work station Safety The COUNT Series contains a high voltage power supply Users may be injured if the case has been opened All internal settings are pre set there are no user adjustments Units that appear defective or have suffered mechanical damage should not be used because of possible electrical shorting of the high voltag...

Страница 11: ...t and documentation Of course we also appreciate good news If you have obtained exciting results with one of our systems please let us know and where appropriate mention us in your publications Retraction of old devices Waste electrical products must not be disposed of with household waste This equipment should be taken to your local recycling center for safe treatment WEEE Reg No De96457402 Produ...

Отзывы: