LxT Manual
Measurement Settings Tabs
4-7
Dosimeter 1 and 2 Tabs
Figure 4-6 shows the
Dosimeter 1
tab.
FIGURE 4-6 Dosimeter 1
The
Dosimeter 1
and
Dosimeter 2
tabs are provided to
permit the evaluation of two independent noise dose data
sets. Other than being on separate tabs, they are identical.
Predefined Setups
In most cases, measurements of this type are setup to
conform to specific standards. The LxT permits the user to
create such setups in a single step by simply selecting the
applicable standard. The standards addressed by the LxT and
the corresponding parameters are as shown in Table 4-1
"Predefined Noise Dosimeter Setups".
Standard
Exchange
Rate
Threshold
Criterion
Level
Hours
OSHA-1
5
90
90
8
OSHA-2
5
80
90
8
ACGIH
3
80
85
8
NIOSH
3
80
85
8
IEC
3
Not Enabled 85
8
Table 4-1 Predefined Noise Dosimeter Setups
Содержание SoundExpert LxT
Страница 1: ...LxT Manual for SoundTrack LxT SoundExpert LxT TM...
Страница 55: ...4 18 Measurement Settings Tabs LxT Manual...
Страница 81: ...5 26 View Spectrum Normalized Optional LxT Manual...
Страница 93: ...6 12 Data Storage After Improper Shutdown LxT Manual...
Страница 110: ...LxT Manual Certification 7 17...
Страница 143: ...11 6 Control Panel Data Explorer LxT Manual...
Страница 167: ...12 24 Options LxT Manual...
Страница 179: ...14 4 User LxT Manual FIGURE 14 5 shows the User tab FIGURE 14 5 User Tab...
Страница 185: ...15 6 File System LxT Manual...
Страница 191: ...16 6 Miscellaneous Parameters LxT Manual...
Страница 206: ...LxT Manual Technical Specifications A 9 Typical Z Weight Frequency Response...
Страница 259: ...A 62 Technical Specifications LxT Manual...