Conductivity set up
32
●
P 1.3 calibration mode setup
1. Press the
key,
P1 COND
screen appears.
3. Press the
ENT
key,
P1.1 CELL
screen appears.
4. Press the
key,
P1.2 UNIT
screen appears.
5. Press the
key,
P1.3 A.CAL
appears.
6. Press the
ENT
key,
A.CAL ON
screen appears with
ON
as default setup.
7. Use the
keys to change the setup to
OFF
. This enables the manual calibration
mode.
8. Press the
ENT
key.
P1.3 A.CAL
screen appears. This indicates completion of
auto calibration mode setup.