CLIENTS user guide SA1000
Page 20
Figure 12. Immunity testing connections
5.3.1 Field probe and door interlock connections
The 8 pin DIN connector on the nose of the cell must be connected to the control system for
immunity testing if using the RF2000. The wiring information is shown in the diagram below.
If using the Laplace RF2000 synthesiser, an 8 way cable is included with the system, otherwise a
mating connector is supplied.
Pin number
Function
1
Hi - Field strength feedback (0
…
2.5v)
2
Lo - Field strength feedback. (0v)
3
Door interlock (closed when door open)
4
0V
5
Not used
6
Range control, pull down for Lo range
7
High range selected (acknowledgement signal)
8
Low range selected (acknowledgement signal)
5.3.2 Connectors
RF
…
N type, 50R
Control cable:
Binder, series 680, metal body
DC power in:
2.1mm DC power jack