6: Network Settings
EDS-MD® Medical Device Server User Guide
40
To Configure Network 1 (eth0) Interface Settings
Using Web Manager
To modify Ethernet (eth0) settings, click
Network
on the menu and select
Network 1 ->
Interface -> Configuration
.
Using the CLI
To enter the eth0 command level:
enable -> config -> if 1
Using XML
Include in your file:
<configgroup name="interface" instance="eth0">
Network 1 (eth0) Link Settings
Physical link parameters can be configured for an Ethernet (eth0) Network Link (see
).
Table 6-2 Network 1 (eth0) Link Settings
Notes:
When speed is
Auto
, duplex must be
Auto
or
Half
.
When speed is not
Auto
, duplex must be
Half
or
Full
.
Fixed speed Full duplex will produce errors connected to Auto, due to duplex
mismatch.
To View or Configure Link 1(eth0) Settings
Using Web Manager
To view Ethernet (eth0) Status, click
Network
on the menu and select
Network 1 -> Link
.
Using the CLI
To enter the eth0 command level:
enable -> config -> if 1
Network 1 Ethernet (eth0)
Link Settings
Description
Speed
Select the Ethernet link speed. (Default is Auto)
Auto
= Auto-negotiation of Link Speed
10 Mbps
= Force 10 Mbps
100 Mbps
= Force 100 Mbps
1000 Mbps
= Force 1000 Mbps
Duplex
Select the Ethernet link duplex mode. (Default is Auto)
Auto
= Auto-negotiation of Link Duplex
Half
= Force Half Duplex
Full
= Force Full Duplex