14
2. Control System
Control Unit:
Microcontroller with internal program memory
Program memory:
32K flash ROM, 2K RAM memory
HMI:
LCD display with LED backlight, soft keypad
Single/double control:
Single or double breast pumping possible
Selectable pumping style mode:
3 pumping style modes are selectable and
controlled from keypad
Adjustable suction level:
8 suction levels are selectable and controlled from pump
keypad and Lansinoh® Baby app when paired and connected.
Suction vacuum level (Expression/Double pumping):
Minimum 100 mmHg,
Maximum 280mmHg.
3. Technical Data Bluetooth version 4.2
Range:
Up to 10m/32.8 ft
4. Power Requirements
AC Adapter (included):
AC Adapter – North American Plug Configuration
100-240VAC wall plug type
AC adapter, galvanically isolated adapter with 9VDC output voltage. The expected
use-life of the pump is 500 hours
For products with an input ground pin, locate near an area accessible to a properly
grounded outlet receptacle.
SYMBOLS
MEANING OF THE SYMBOLS
Follow instructions for use
Identifies the manufacturer
Indicates the date of manufacture
Type BF applied part
Class II equipment
Keep the device dry
Keep the device away from sunlight
Transport/
Storage
Operation
41°F
95°F
41°F
95°F
Indicates the temperature limitation for operation, transport and storage.
Transport/
Storage
20%
95%
Operation
20%
80%
Indicates the humidity limitation for operation, transport and storage.