11
5. Operation
ENGLISH
5.3.3. Two- and three-point calibration
Follow the diagram below.
The equipment will repeat the process described for the first
buffer solution with the second and third.
Stir gently holding the sensor by the handle. Do not hold it by
the tube to avoid heating of the buffer solution.
Notes:
if the measuring instrument is not used for 5 minutes, it
shuts off automatically.
If errors occur during calibration, an error message will appear
on screen (see pg. 19).
When you switch buffers, rinse the electrode with deionized water.
The buffer solutions in the calibration tubes can be used for
approximately four or five calibrations. It is recommended to
use fresh buffers.
Screw the
electrode
to the 1st tube.
Screw the
electrode
to the 2nd tube.
To exit
calibration
(One point
calibration)
PH1_HACH_LANGE_0710.indd ANG:11
PH1_HACH_LANGE_0710.indd ANG:11
7/9/10 15:40:49
7/9/10 15:40:49
Содержание sension+ ph1
Страница 2: ...sensION PH1_HACH_LANGE_0710 indd 2 sensION PH1_HACH_LANGE_0710 indd 2 7 9 10 15 40 30 7 9 10 15 40 30...
Страница 4: ...sensION PH1_HACH_LANGE_0710 indd 4 sensION PH1_HACH_LANGE_0710 indd 4 7 9 10 15 40 31 7 9 10 15 40 31...
Страница 28: ...sensION PH1_HACH_LANGE_0710 indd ALE 2 sensION PH1_HACH_LANGE_0710 indd ALE 2 7 9 10 15 41 04 7 9 10 15 41 04...
Страница 50: ...sensION PH1_HACH_LANGE_0710 indd FRA 2 sensION PH1_HACH_LANGE_0710 indd FRA 2 7 9 10 15 41 37 7 9 10 15 41 37...
Страница 73: ...sensION PH1_HACH_LANGE_0710 indd CON 3 sensION PH1_HACH_LANGE_0710 indd CON 3 7 9 10 15 42 10 7 9 10 15 42 10...
Страница 74: ...sensION PH1_HACH_LANGE_0710 indd CON 4 sensION PH1_HACH_LANGE_0710 indd CON 4 7 9 10 15 42 10 7 9 10 15 42 10...