Landtek PH-ORP 2000 Скачать руководство пользователя страница 3

www.landtek.com.cn 

www.landteknet.com

7.1 Clean the ORP electrode with de-ionized or 
      distilled water.
7.2 Place the electrode in the mv buffer solution. 
7.3 When display stabilizing, check if the 
      reading on the display is close to the standard 
      value of mv solution. If the error lies in 

±

35

      mv. If yes, indicates the electrode tested can 
      still be used. If not, replace it by new one.

The  High  setting  determines  the  operating 
parameters  for    Relay  High;  while  low  setting 
defines the operating parameters for Relay Low. 

8.HIGH AND LOW SETTINGS FOR PH 

HiAL on the upper section of the Display. It 
takes about 3 second from pressing the M key. 
Then Press the up key      or down key      to 
adjust the digit on the auxiliary section to the 
desired value which controls the high relay 
output.

section. Input the hysteresis value (dead 
band) on the auxiliary section by Pressing the 
up key     or down key     . e.g. to control the PH 
value between 9-10 by adding acid, just set 
HiAL to 

10

, and hb to

1

section. Then Press the up key      or down key   
       to adjust the digit on the auxiliary section 
to the desired value which controls the low 
relay output.

8.1 Press the M key and not release it until the 

8.2 Press the M key and hb shows on the main 

8.3 Press the M key and LoAL shows on the main 

CONTROLLER

6

Carefully unpack the instrument and accessories. 
Inspect  for  any  damages  made  in  shipment.  The 
Controller shall be installed and operated only in 
the  manner  specified  in  the  Instruction  manual. 
Only skilled, trained or authorized person should 
carry out installation, setup and operation of the

4.ASSEMBLY AND INSTALLATION

 instrument.
Please follow steps below.

A.Installation of composite electrode.
b.Installation of PH/ORP controller
c.Connect the probe to the controller by an 
   appropriate cable.
d.Connect the pump or electromagnetic valve
e.Power on the controller
f.Calibrate the controller using the standard 
   sample.
g.Set the control limits

PH/ORP

A/PH

R

P

A/R

A typical system is shown bellow.

A/PH:  a Relay contact inside of the PH/ORP 
             controller
A/R:    a contact of the power relay

3

R:  a power replay
P:  pump or valve

M (3 s)

M

Hb/0.00

LoAL/0.00

Lb/0.00

CS/0

CH/20.00

Fun/PH 

CAL/...... 

Display 

HiAL/0.00

M

M

M

M

M

M

Low limit, adjusted by the up key      or 

  

down

Hysteresis value (Dead Band) for High limit,
adjusted by the up key     or down key   

High limit, adjusted by the up key      or 

 

down

Hysteresis value (Dead Band) for Low  limit,
adjusted by the up key       or down key  

Set value for manual temp. Compensation
adjusted by the up key      or down key     

Indication of current use: PH or ORP
changed by the up key      or down key      

2-point calibration
please refer to Calibration part for detail.

Status

  key     

 Key  

=0: without temperature compensation 
=1: 
=2: 

manual temperature compensation
automatic temperature compensation

Auxiliary value

5. MENU OPERATION DIAGRAM

6.PH CALIBRATION

6.1 Clean the pH electrode and temperature probe   
  
6.2 Place the electrode in the first buffer(PH6.86) 
      solution. In ATC mode, you must also immerse
 

with de-ionized or distilled water.

4

M

  temperature  probe  in  the  buffer  solution. 

just press the M key and not release it until the 
HiAL on the upper part of the Display. It takes 
about 3 second from pressing the M key. Then 
press  and  release  the  M  key  till 

CAL

  on 

the main section and  ----  on the auxiliary

the

6.3 When the reading on the display is stabilizing, 

 section.

6.4 Press the up key    or down key

 to select P1 

      shows on the auxiliary section.
6.5 
      
       adjust to 6.86 by the up key     or down key      .
6.6 Press      key and  6.86  shows on the main 
       
      section.
6.7 Press the up key     or down     key to select 
      yes  shows on the auxiliary section. 
6.8 Press the M key and P1 shows on both main 
      and auxiliary sections.
6.9 After a second,  CAL  shows on the main 
      section and  ----  on the auxiliary section. 
      above procedures makes P1 calibration ok.
6.10 Repeat steps from 6.1 to 6.9, just change the   
     PH buffer (standard) to PH4.00 (or 9.18),  and 
     P2 replacing P1 to complete the P2 calibration.

    

 And 6.86 on the auxiliary section.  if not 6.86, 

section  and  no  shows  on  the   auxiliary

7.ORP CALIBRATION

      To calibrate ORP electrode does not need PH   
      buffer instead of mv buffer to test if the ORP 
      electrode is still workable.

5

M

Press the    key, and P1 shows on the main section 

Отзывы: