To check whether the tester is accurate, just 
Insert  the  indenter  into  the  hole  of  the 
calibrated  test    Block.   Apply  enough  force 
to  make  firm  contact  between  the  top 
surface  of  the  test    block  and  the    presser   
foot.    The  reading  should  agree  with  the 
value  stamped  on  the  test  block.  If  not,  just 
carry  out  Zero  calibration  and  High  end 
calibration. 

7.1 Zero calibration 
      Hold the durometer vertically with the point 

of  the  indenter  hanging  in  the  air,  the 
reading on the display should be “0”. If not, 
depress  the  “Zero  key”  to  make  the  tester 
display “0”.

7.2 High end calibration

Just  place  the  indenter  onto  a  flat  glass, 
apply  enough  force  to  make  firm  contact 
between  the  glass  and  the  presser  foot.  The 
readings  on  the  display  should  lie  between 
99.5  and  101.  If  not,  press  “CAL  key”  to 
carry out high end calibration. 

8.   BATTERY REPLACEMENT 

  

8.1  When  the  battery  symbol  appears  on  the 

display, it is time to replace the batteries. 

8.2 Slide the Battery Cover away from the tester 

and remove the batteries. 

8.3  Install  batteries  paying  careful  attention  to 

polarity. 