BIOGAS 5000 Gas Analyzer
OMBG5KN4.7.1
Page
53
of
61
9.8
Restore to factory settings
This option will reset the gas analyser to all of its factory programmed calibration
settings and will clear ALL the user defined calibration points. It will not affect or
remove ID’s or readings from the analyzer.
1)
Select ‘Key 2 - Restore to factory’ followed by the soft-key ‘Confirm’ or ‘Cancel’.
2)
A validation message is displayed ‘Reset user calibration?’ Press the soft-key
‘Confirm’ to continue with the factory settings or soft-key ‘Cancel’ to cancel the
operation and return to the Gas Check menu.
9.9
Calibration history
The BIOGAS 5000 gas analyzer logs user calibrations in the ‘History’ application. This
can be used as an aid to ensuring that gas measurements are valid and accurate. Both
good and failed calibration results are recorded for each channel calibrated.
1)
Select ‘Key 3 – History’.
2)
The operator may view the calibration data stored. Use the soft-key ‘Filter’ to add a
sort filter to the history enquiry.
9.10
Calibration summary
The BIOGAS 5000 gas analyzer has the facility to log the history of user calibrations.
1)
Select ‘Key 4 – Summary’.
2)
The operator may view the calibration data history stored by ID, technician,
timestamp, type and calibration result. Use the soft-key ‘Exit’ to exit and return to
the ‘Gas Check’ menu.