FILLING: SINGLE-USE DISK
TESTING
QUICK
START GUIDE
BEFORE FIRST USE
¬
Fully charge battery.
¬
Ensure your Spin Touch has the latest fi rmware:
1. Download and install the WaterLink Connect Desktop application for Windows
®
at
www.waterlinkconnect.com/Downloads/Application
.
2. Plug meter into the computer with the provided USB cable and launch WaterLink
Connect desktop application from the Start Menu.
3. Any fi rmware updates will automatically be applied to the meter.
1. Fill the disk using slow, even pressure. The sample
water will fi ll the spaces between the baffl
es in a
counterclockwise order. Each space will fi ll from the
bottom to the top. Sample water should be added until
the sample water in the fourth chamber fi lls
to the top
of the chamber slightly past the embossed fi ll line. It is
OK to fi ll slightly past the fi ll line.
Correct Fill
(Fill line)
2. Do not overfi ll the disk. If the disk is overfi lled, sample
water will fl ow out of the overfl ow hole in the center of
the disk. The disk is not leaking. Dry the disk and run
the test.
Over
Filled
(Dry disk before placing in meter)
3. Do not under fi ll the disk. If the disk is under fi lled, the
reagent chambers will not fi ll entirely and results will
be inaccurate.
Under
Filled
4. Wet disks should be dried thoroughly with a lint free
wipe. The disk should be handled by the edges.
5. Disks should be fi lled and used within 10 minutes. They cannot be fi lled ahead of time.
1. Press
and hold until the meter turns
on.
2. Tap . Select a sanitizer type. Tap
to confi rm.
3. Tap . Select a disk series (found on
disk packaging). Tap to confi rm.
4. Remove disk from packaging.
5. Use syringe (1189) to fi ll disk with
pool or spa water sample.
6. Insert disk. Cover disk with Universal
Disk Cover. Close lid.
7. Tap to start test. Tap
to stop the test.
8. Results will be displayed.
9. Select an option.
- Tap to return to the Test Screen.
- Tap highlighted
to save test results
to test log if Auto Save is not enabled.
- Tap the hightlighted
to send the results
to the enabled Mobile Bluetooth Printer.
- Tap highlighted
to send results to
Bluetooth enabled device.
10. Press and hold
for 2 seconds to turn
meter off .
NOTE: For water samples over 100 °F (38 °C) subtract 0.3 from pH result or, for the most accurate result, wait
until water sample is below 90 °F (32 °C) to test.