7. Tap

to start test.

8.  Results will be displayed.

9.  Compare the results on the display to the values on the chart shown on the pouch. If 

the results are not within the range shown on the pouch, contact Support.

10. Replace the Meter Check Disk in the foil pouch for storage.

Alignment Procedure

The Meter Check Disk is inserted in the meter and Calibration is chosen from the Settings 
Menu to perform an angle calibration which checks the alignment of the hub and disk. The 
results are analyzed and reported as Pass or Fail. If the measurements pass, the settings 
will be saved. If the analysis fails, contact Support. 

1.  Follow the Cleaning procedure on page 18 to clean the light chamber and optic lenses. 

2. Tap   to go to Settings.

3. Tap 

Calibration

.

4.  Remove the Meter Check Disk from the foil pouch. DO NOT remove the black cover 

from the disk.

5.  Insert the Meter Check Disk. Close the lid.

6. Tap 

Start

to begin the meter check procedure.

7.  If the meter is performing satisfactorily, “Angle Calibration Successful” will be 

displayed and the settings will be saved. If the meter is not performing satisfactorily, 
“Angle Calibration Unsuccessful. Contact Tech Support.” will be displayed.

8. Tap

g

to return to the Testing Menu and resume testing.

email: [email protected]

phone: 800-344-3100 option 3 | Mon-Fri 9am-5pm EST

HELPFUL HINTS

¬

Do not touch top or bottom of disk. Handle disk by the edge.

¬

Do not fi ll disk while in the meter. Fill disk on clean, dry surface.

¬

Fill the disk on a dark surface to more easily see the sample water.

¬

The disk should not contain any large air bubbles. Air bubbles will result in erroneous 
results.

¬

Always use the disk cover.

¬

Only the Universal Disk Cover (Code 1719) can be used with the WaterLink

®

 Spin 

Touch

®

.

¬

Empty syringe of old sample before fi lling with next sample.

¬

Remove fi lled disk from meter after testing. Do not travel with fi lled disks in meter. 
They may leak.

¬

Keep the chamber clean and dry. Gently swab LED and photodiode lenses located 
around the hub with a cotton swab dampened with streak-free window cleaner. Do 
not use alcohol. It will leave a thin fi lm over the lenses when dry.

Содержание WaterLink Spin Touch FF

Страница 1: ...WaterLink n F Touch F Code 3587 Instruction Manual...

Страница 2: ...will require updates To do so 1 Visit http softwarecenter lamotte com and download the WaterLink Connect application for Windows 2 Plug the Spin Touch into the computer using the included USB cable 3...

Страница 3: ...be transferred from the meter into the application Mobile apps for LaMotte software products are available from iTunes for iOS devices and Google Play for Android devices Learn more about LaMotte sof...

Страница 4: ...2 Do not over ll the disk If the disk is over lled sample water will ow out of the over ow hole in the center of the disk The disk is not leaking Dry the disk and run the test Overfilled Dry disk bef...

Страница 5: ...in progress and the disk is spinning Do not open the lid when the disk is spinning Care should be taken when closing the lid Do not slam the lid Wiring between the lid and the body of the photometer p...

Страница 6: ...test results to the test log if Auto Save is not enabled Tap the hightlighted to send the results to the enabled Mobile Bluetooth Printer Tap the highlighted to send the results to a Bluetooth enabled...

Страница 7: ...le at a time Auto Save Tests must be disabled If Auto Save Tests is disabled will be highlighted on the Test Results screen After the test has been run tap to save the results for that water sample to...

Страница 8: ...nge ts into the ll hole on the disk The syringe tip should not be removed from the syringe Syringes should be cleaned between water samples Pump air in and out of the syringe a few times to clear the...

Страница 9: ...ch to a Windows based PC When used with the AC Power Adapter it connects the meter to an AC outlet WARNING only use the USB cable and wall adapter that are supplied with the kit Make no substitutions...

Страница 10: ...ment of the hub and disk Tap to exit to the Settings menu Power Options There are three power options Auto Dim Time Auto O Time and Power Tap the options then tap a selection Tap selection Tap to exit...

Страница 11: ...f range of the reagent system wil be RED RED test results may not be accurate Go SETTINGS Other Settings to turn the Ranges Enabled feature OFF Testing samples for Nitrate immediately after treatment...

Страница 12: ...ange pH Blank Low Range Hardness High Range pH Empty Nitrate Phosphate Ammonia Fish Farming Salt Water Disk FF203 Code 4352 pH Phosphate Calcium Magnesium Alkalinity Nitrite Nitrate Ammonia pH Nitrite...

Страница 13: ...erLink Spin Touch FF reagent disks see page 12 SPECIFICATIONS Instrument Type Centrifugal Fluidics Photometer Wavelengths interference lters 390 nm 428 nm 470 nm 525 nm 568 nm 635 nm Display Color Cap...

Страница 14: ...ter Resistance Rubber over molded base rubber USB Port Plug gasketed display and hinge Electrical Rating Rated voltage 5V Rated power of input current 1 6 A at USB C Auto O Yes default 15 only with ba...

Страница 15: ...l disk correctly see FILLING Range Error Raw data out of range Contact Support Output Error Decreased light intensity Possibly dirty lens Clean lens see CLEANING Follow Range Check Procedure If error...

Страница 16: ...r not on The Spin Touch will only print to the Mobile Bluetooth Printer Code 5 0066 Turn printer on Connect to Mobile Bluetooth Printer Code 5 0066 TROUBLESHOOTING WITH THE METER CHECK DISK Do not att...

Страница 17: ...performing satisfactorily Angle Calibration Successful will be displayed and the settings will be saved If the meter is not performing satisfactorily Angle Calibration Unsuccessful Contact Tech Suppor...

Страница 18: ...ber must be obtained from LaMotte Company by calling 800 344 3100 ext 3 US only or 410 778 3100 ext 3 faxing 410 778 6394 or emailing softwaresupport lamotte com Often a problem can be resolved over t...

Страница 19: ...nts is available 24 hours a day from the National Poison Control Center at 1 800 222 1222 or by contacting the 24 hour emergency line for ChemTel at 1 800 255 3924 USA Canada Puerto Rico For locations...

Страница 20: ...defects resulting from action of the user such as misuse improper wiring operation outside of speci cation improper maintenance or repair or unauthorized modi cation LaMotte Company speci cally discla...

Отзывы: