background image

LaMotte TRACER Fluoride PockeTester

TM

 1756 Instruction 

5

CONTENTS

Fluoride TRACER Kit

Code 1756

Includes:
  Fluoride TRACER Body
  Flat Surface Fluoride Electrode

 TISAB TesTabs (20)

  Tablet Crusher
  Sample Cup
  Lanyard
  Protective Sensor Cap
  Batteries, 3V (4)

PARTS & ACCESSORIES

TISAB TesTabs (100)

Code 7024-J

Tablet Crusher

Code 0175

Flat Surface Fluoride Electrode

Code 1757

Fluoride Standarad, 1.0 ppm, 3800 mL

Code 2798-M

METER DESCRIPTION

Front Panel Description

1. Battery compartment cover
2. LCD Display
3. MODE/HOLD key
4. CAL/RECALL key
5. ON/OFF key
6. Electrode Retaining Collar (ring)
7. Electrode Sensor

(Note: The Electrode storage cap is not shown)

Содержание Tracer PockeTester 1756

Страница 1: ...Water Testing Leader Since 1919 Tracer Fluoride PockeTesterTM Code 1756 99 Washington Street Melrose MA 02176 Phone 781 665 1400 Toll Free 1 800 517 8431 Visit us at www TestEquipmentDepot com...

Страница 2: ...WARNING This set contains chemicals that may be harmful if misued Read cautions on individual containers carefully Not to be used by children except under adult supervision...

Страница 3: ...ation 7 Calibration 7 Measurement 8 Storage and Maintenance 8 Operation Preparation for use 8 Powering the Fluoride TRACER 9 Power On Diagnostic 9 Calibration 9 Calibration Frequency 10 Slope Adjustme...

Страница 4: ...s Direct reading of ppm units Direct reading of relative mV units Automatic shut down after 12 minutes to preserve battery life Internal error detection Convenient TISAB TesTab reagent SPECIFICATIONS...

Страница 5: ...ve Sensor Cap Batteries 3V 4 PARTS ACCESSORIES TISAB TesTabs 100 Code 7024 J Tablet Crusher Code 0175 Flat Surface Fluoride Electrode Code 1757 Fluoride Standarad 1 0 ppm 3800 mL Code 2798 M METER DES...

Страница 6: ...eference electrode and temperature measurement system The high resistance electrode signals are impedance converted to a low resistance output in the sensing module to ensure stable and noise free per...

Страница 7: ...t with Tablet Crusher 0175 and mix until the tablet disintegrates Or prepare a traditional 1 0 ppm f uoride standard and TISAB reagent or use premade mixed TISAB and 1 0 ppm standard Pour 20 mLs of th...

Страница 8: ...by pressing the MODE HOLD key for approximately 3 seconds Storage and Maintenance 1 After use store the electrode in an analyzed sample Fluoride standards plus TISAB tablet 2 Thoroughly wipe the senso...

Страница 9: ...has been chosen 1 Fill a sample cup to 20 mL with a 1 ppm f uoride standard Add one TISAB TesTab 7024 Crush tablet with Tablet Crusher 0175 and mix until the tablet disintegrates Or prepare a traditio...

Страница 10: ...or 10 0 ppm Measurements 1 Prepare an unknown solution by adding crushing and dissolving one TISAB TesTab 7024 in 20 mL of the sample or by adding TISAB reagent to the sample in the same dilution rati...

Страница 11: ...e a reading Release the key when the memory location number appears on the lower display 2 After approximately 30 seconds measurement duration the meter will return to the HOLD mode and another readin...

Страница 12: ...ys on the electrode and the main body housing and then carefully push the electrode into the meter socket until it is fully seated CAUTION Take care to align pins carefully Bent or broken pins will ca...

Страница 13: ...in transit This warranty does not apply to defects resulting from action of the user such as misuse improper wiring operation outside of specif cation improper maintenance or repair or unauthorized mo...

Страница 14: ...LaMOTTE COMPANY Helping People Solve Analytical Challenges 1756 MN 12 13 Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...

Отзывы: