3.3.2 0% Calibration
1. Rinse the probe thoroughly with DI
water or rinse solution.
2. Press
MODE
to select % saturation.
3. Dip the probe into a zero oxygen
solution. Stir gently with the probe. Wait
for the reading to stabilize.
4. Press
CAL
. The CAL indicator and
intended percentage calibration point
(0%) will display brief y, before CAL
f ashes. The % value based on the
factory default calibration will be shown.
5. Press
HOLD/ENTER
to conf rm the
calibration. The meter will display
“
CO
”, automatically calibrate to
0.0% saturation, and then return to the
measurement mode.
A value with respect to the
factory default calibration
will be displayed
%
CAL
A
%
HOLD
ENTER
CAL
CAL
%
Calibration point will be
displayed for 1.5 seconds
%
“CO” will displayed for
1.5 seconds