* Sulfide Reagent A: Code V-4458LWT-G
* Sulfide Reagent B: Code V-4459-E

Sulfide Reagent C: Code 4460LWT-G

SULFIDE

ColorQ Range: 0.0-3.0 ppm

Color

Q

5

6

4. Add 3 drops of

.

SULFIDE REAGENT B

7. Cap. Invert 3 times to mix. Insert tube into ColorQ.

3. Add 5 drops of

.

SULFIDE REAGENT A

5. Cap and invert 3 times to mix. Allow tube to sit for ONE minute.

2. Fill a clean tube to the 5 mL line with water sample.

6. Add 5 drops of

.

SULFIDE REAGENT C

8. Press button to read “SuL” Sulfide in ppm. Remove tube.

1. Press button to go to “SuL” Sulfide.

DILUTION

Samples with sulfide concentrations above 3 ppm must be diluted. The
meter will display “Hi”.
A. Use the 1 mL pipet to add 1 mL water sample to a test tube.

C. Blank with the diluted sample from Step B.  Follow the test procedure

with the diluted sample from Step B.  Multiply the displayed result by 5.

D. Follow the test procedure with this diluted sample, starting with step 3.

B. Dilute to the 5 mL line with distilled or deionized water. Cap and mix.

E. Multiply display result by 5.

Содержание ColorQ DW

Страница 1: ...own Maryland 21620 USA bAt Battery is low Use only LaMotte replacement reagents If the unit should auto off during testing rinse and fill one tube to Blank and return to testing Insert tubes into Colo...

Страница 2: ...s button to Blank the meter and go to pH Remove tube 1 Fill sample bottle with water sample Replace cap 2 Fill clean tube to the 5mL line with water sample 3 Insert tube into ColorQ as shown pH 2 pH C...

Страница 3: ...ardness in gpg Remove tube 4 Cap Invert 3 times to mix Insert tube into ColorQ DILUTION A Use the 1 mL pipet to add 2 mL water sample to a clean test tube C Blank with the diluted sample from step B E...

Страница 4: ...etely Insert tube into Color Q DILUTION Samples with iron concentrations above 3 ppm must be diluted The meter will display Hi A Use the 1 mL pipet to add 1 mL water sample to a test tube B Dilute to...

Страница 5: ...ube 1 Press button to go to FCL Free Chlorine 3 Add 1 DPD 1 Tablet Cap Shake for TEN seconds Invert slowly 5 times 2 Fill a clean tube to the 5 mL line with water sample Free Chlorine and Total Chlori...

Страница 6: ...Add 5 drops of SULFIDE REAGENT C 8 Press button to read SuL Sulfide in ppm Remove tube 1 Press button to go to SuL Sulfide DILUTION Samples with sulfide concentrations above 3 ppm must be diluted The...

Страница 7: ...trogen in ppm Remove tube 4 Cap Invert 30 times per minute for TWO minutes 3 Add 1 Tablet Crush with tablet crusher Immediately slide tube into Protective Sleeve NITRATE NOTE Nitrate Tablets are sensi...

Страница 8: ...ness results Er 3 Insufficient light is reaching the detector Sample may be too dark or tube may be in sideways Turn off meter and retest Er 7 The blank was not properly set Turn off meter and retest...

Отзывы: