12. Press
*
I
OK
to select
Scan Sample
.
T u r b i d i t y
( F )
* S c a n S a m p l e
1 6 : 0 2 : 1 9
0 1 / 0 4 / 0 5
13. Observe the result.
T u r b i d i t y
( F )
0 . 5 4
N T U
* S c a n S a m p l e
q
1 6 : 0 2 : 1 9
0 1 / 0 4 / 0 5
14. Press
and then press
*
I
OK
to
select
Calibrate
.
T u r b i d i t y
( F )
0 . 5 4
N T U
* C a l i b r a t e
q
1 6 : 0 2 : 1 9
0 1 / 0 4 / 0 5
15. Use the
or
to change the
highlighted digits on the display to
match the concentration of the
turbidity standard. Press
*
I
OK
to
accept a digit and move to the
next digit.
C a l i b r a t e
0 0 . 5 4
q
,
Q
,
p
1 6 : 0 2 : 1 9
0 1 / 0 4 / 0 5
C a l i b r a t e
0 0 . 5 4
q
,
Q
,
p
1 6 : 0 2 : 1 9
0 1 / 0 4 / 0 5
C a l i b r a t e
0 0 . 5 4
q
,
Q
,
p
1 6 : 0 2 : 1 9
0 1 / 0 4 / 0 5
C a l i b r a t e
0 0 . 5 0
q
,
Q
,
p
1 6 : 0 2 : 1 9
0 1 / 0 4 / 0 5
ANALYSIS & CALIBRATION
27
0
4
5
0
0