KEEP THESE INSTRUCTIONS

METER CHECK DISK  |  Code 1705-EU

TROUBLESHOOTING WITH THE  

METER CHECK DISK

IMPORTANT!!  Do not attempt to separate 

the components of the Meter Check Disk 

(Code 1705- EU). The Meter Check Disk 

consists of a disk with a permanently 

attached cover. Do not fill the Meter Check 

Disk with water. Water is not used in the 

Meter Check Disk.
CALIBRATION OPERATIONS

  The Meter 

Check Disk (MCD) is used to perform two 

calibration operations – Check Calibration 

and Start Calibration. The Start Calibration 

procedure should only be performed if the 

meter fails the Check Calibration procedure.

Check Calibration 

 Meters are calibrated 

at the time of manufacture. However, it 

is possible for the calibration settings to 

be lost due to power anomalies or other 

circumstances. The Meter Check Disk is 

used in the Check Calibration procedure to 

determine whether the alignment of the 

hub and disk are correct. For some meters, 

it also evaluates the brightness of the 

individual LEDs. 
1. Follow the CLEANING procedure to clean 

the light chamber and optic lenses.

2. From the main test page, tap   to select 

Settings.

3. Tap 

Calibration

.

4. Tap 

Check Calibration

.

5. Remove the Meter Check Disk (Code 

1705-EU) from the foil pouch. DO NOT 

remove the black cover from the disk. 

Insert the Meter Check Disk (MCD) into 

the meter and close the lid.

6. Tap 

Start

.

7. The meter will briefly spin. When it is 

complete, six channel values will be 

displayed. Compare the displayed channel 

values to those printed on the MCD pouch. 

If the displayed channel values are within 

the ranges provided on the Meter Check 

Disk pouch, the meter is calibrated and 

performing normally. If the displayed 

channel values are not within the ranges 

provided on the Meter Check Disk pouch, 

perform the Start Calibration procedure. 

Note:

 Range specifications are specific to 

the disk identified by the serial number on 

the pouch. The range specifications will 

vary from disk to disk. The exact readings 

from a specific disk may vary from meter 

to meter. 

8. Tap   to return to the test screen.
9. Remove the MCD from the meter and 

return to the foil pouch for storage.

Start Calibration 

 The Start Calibration 

procedure calibrates the alignment of 

the hub and disk for all meters and sets 

the brightness of the individual LEDs for 

compatible meters. Before performing 

this calibration procedure, run the Check 

Calibration procedure to determine whether 

the meter is calibrated and operating 

normally. The Start Calibration procedure 

should only be completed if the meter fails 

the Check Calibration procedure.
1. Follow the CLEANING procedure to clean 

the light chamber and optic lenses.

2. From the main test screen, tap   to 

select Settings.

3. Tap 

Calibration

.

4. Tap 

Start Calibration

.

5. Remove the Meter Check Disk (Code 

1705-EU) from the foil pouch. DO NOT 

remove the black cover from the disk. 

Insert the Meter Check Disk. Close the lid.

6. Tap 

Start

.

7. When the calibration is complete the 

message “Angle Calibration Successful” 

will appear. Meters that can also perform 

an LED calibration will display “LED 

Calibration Successful”.

8. Tap   to return to the test screen.
Angle Calibration checks the alignment of 

the hub and disk. The LED Calibration sets 

the brightness of the individual LEDs. The 

results are reported as pass or fail. If the 

measurements pass, the settings will be 

saved, and the meter is calibrated. If the 

analysis fails, contact Support. 

Storage

  Store the disk in a cool, dry place in 

the foil pouch with the desiccant. The disk is 

sensitive to moisture. Exposure to excessive 

heat or moisture may cause degradation 

of the disk. The Meter Check Disk contains 

delicate materials. Do not drop, jar, or twist 

the disk. Do not get the disk wet. Keep the 

Meter Check Disk clean and dry.

Содержание 1705-EU

Страница 1: ...Range specifications are specific to the disk identified by the serial number on the pouch The range specifications will vary from disk to disk The exact readings from a specific disk may vary from me...

Страница 2: ...impresos en el estuche del MCD Si los valores de los canales mostrados se encuentran dentro de los rangos proporcionados en el estuche del Disco de comprobaci n del medidor el fot metro est calibrado...

Страница 3: ...ce ou d autres circonstances Le disque de v rification du compteur est utilis dans la proc dure de v rification de l talonnage pour d terminer si l alignement de l axe et du disque est correct Pour ce...

Страница 4: ...pas tomber ne le secouez pas et ne le tordez pas Ne mettez pas le disque en contact avec de l eau Conservez le disque de v rification de photom tre propre et sec GUARDEESTASINSTRU ES DISCODEVERIFICA...

Страница 5: ...Angle Calibration Successful Os medidores que tamb m podem efetuar uma calibra o do LED exibir o Calibra o com xito LED Calibration Successful 8 Toque em para voltar ao ecr de teste A Calibra o do ng...

Страница 6: ...LED Prima di eseguire tale procedura completare quella di verifica della calibrazione per determinare se il fotometro calibrato e funziona con normalit La procedura di avvio della calibrazione deve e...

Страница 7: ...angezeigten Kanalwerte mit den aufgedruckten Werten auf dem Beutel der Messger t Pr fscheibe Wenn sich die Kanalwerte innerhalb der auf dem Beutel der Messger t Pr fscheibe angezeigten Intervalle befi...

Страница 8: ...ibratiecontrole en een Kalibratiestart De procedure voor de Kalibratiestart moet uitsluitend worden uitgevoerd wanneer de meter de Kalibratiecontrole niet doorkomt Kalibratiecontrole Meters worden tij...

Страница 9: ...metercontroleplaatje bevat kwetsbare materialen Laat de schijf niet vallen en kras of verdraai hem niet Laat de schijf niet nat worden Houd de metercontroleplaatje schoon en droog F LJDESSAANVISNINGA...

Страница 10: ...p en sval torr plats i p sen med torkmedlet Skivan r fuktk nslig Exponering f r stark v rme eller fukt kan orsaka nedbrytning av disken M taren Kontrollskiva inneh ller mt ligt material Tappa inte sk...

Страница 11: ...ve optik lensleri temizleyin 2 Ana test ekran ndan se ene ine t klayarak Ayarlar se ene ini belirleyin 3 Kalibrasyon Calibration se ene ine dokunun 4 KalibrasyonBa latma Start Calibration se ene ine...

Страница 12: ...LaMOTTE COMPANY 802 Washington Ave Chestertown Maryland 21620 USA lamotte com Software Support 800 344 3100 x3 softwaresupport lamotte com 2022 LaMotte Company 61705 EU REV1 05 23 2022...

Отзывы: