LAMBDA REDOX measurement unit

 

 

Operation MANUAL

 

 

 

www.lambda-instruments.com

 

 

8

 

6  APPENDIX 

6.1  RS communication protocol for LAMBDA REDOX Potential 

measurement unit 

6.1.1  Communication settings: 

Speed: 

2400 Baud 

Data format: 

8 bit, odd parity, 1 stop bit 

Transmit-Receive switching: 

DTR (usually automatic switching) 

Transmit-Receive interval: 

10 ms 

 

 

 

6.1.2  Data format for transmission from the computer (master): 

# ss mm 

G

 qs c 

send data 

# ss mm 

M

 qs c 

send data 

 

where: 

ss 

is the address of the slave receiver 

mm 

is the address of the transmitter (master) 

M

 or 

 

command to send data 

qs 

control sum 

carriage return CR (ASCII 0D) 

 

 

 

 

 

 

6.1.3  Data format for transmission to the computer (master): 

< mm ss a ddd qs c 

where: 

a

 

– for positive value 

– for negative value

 

ddd

 

data (3 ASCII number 0.......9 - transmission from high to low)

 

qs

 

control sum in hexadecimal format (2 ASCII symbols 0.......9 A B C D E F) 

 

 

 

 

 

 

6.2  How to set the REDOX measurement unit address? 

To look up/modify the instrument address, disconnect the 8-pole cable from the REDOX 
measurement unit. 

Press the 

ADRS

 button continuously and at the same time connect the 8-pole cable to 

REDOX measurement unit again. The message 

“A”

 and two numbers will appear on the 

display. This number from 00 to 99 is the current address of the powder dosing instrument.  

To change the address press the buttons 

Λ Λ Λ

 under the display until the desired number is 

obtained.  

To confirm and save the address, press the 

OK

 button. 

 

 

Содержание REDOX

Страница 1: ...LAMBDA REDOX Red Ox Potential measurement unit OPERATION MANUAL REDOX Measurement Unit...

Страница 2: ...rear side of the MINIFOR base unit LAMBDA Laboratory Instruments is developer and producer of special laboratory instruments mainly for biotechnology microbiology food and agricultural chemical and p...

Страница 3: ...Red Ox ORP potential 3 2 PC control 3 2 1 Setting up the address of REDOX 3 2 2 Configuration of SIAM software for REDOX 4 3 Technical specifications 6 3 1 General specification 6 3 2 Remote control...

Страница 4: ...erred by the 8 pole cable art no 4810 which also functions as a RS connection cable of the MINIFOR When connected with MINIFOR through the cable numbers on the display of LAMBDA REDOX will light and i...

Страница 5: ...2 A new window will pop up to choose the desired configuration file Under the folder CONFIG select the desired CFG file Figure 2 2 3 The selected program will open Creating a new configuration file f...

Страница 6: ...edox click on the display and right click Choose Channel choice from the list Figure 2 2 6 CHANNEL TO VISUALISE window will pop up Under the DEVICE type or browse for redoxm1 Redox Minifor 1 and selec...

Страница 7: ...LAMBDA REDOX Red Ox potential ORP measuring unit for MINIFOR Fermentor Bioreactor Working range 999 to 999 mV in 1 mV steps Interface RS 485 Power supply Powered by MINIFOR Fermentor Bioreactor Dimen...

Страница 8: ...bined redox pH and temp probe 42 cm with Variopin connector Mettler Toledo Control 4810 Pump remote control analog and digital cable 8 poles connector 800202 Quadruple plug box Power and RS connection...

Страница 9: ...0D 6 1 3 Data format for transmission to the computer master mm ss a ddd qs c where a r for positive value l for negative value ddd data 3 ASCII number 0 9 transmission from high to low qs control sum...

Страница 10: ...4 brown 12 V 5 white input remote ON OFF 0V ON 3 12 V OFF this logic can be inversed on demand 6 pink earth ground GND 7 red RS 485 B 8 blue RS 485 A zero line connected to the contact no 3 LAMBDA Lab...

Отзывы: