3.

 

Product Handling 

 

™

 

Do not operate this unit before reading this manual. 

™

 

Before operating Laird Technologies’ recirculating liquid chiller, be certain   

that: 

1.

 

the unit is positioned vertically. 

2.

 

the liquid reservoir and entire recirculating system are filled with 

coolant. 
3.

 

the air inlet and outlet are at least 12” away from any restrictions. 

 

WARNING  

Electrical Shock Hazard 

 

A.

 

DO NOT OPERATE UNIT WITHOUT COOLANT IN THE SYSTEM.

 

B.

 

ENSURE THAT THERE IS A CIRCUIT BREAKER ON THE LINE AND 
THAT IT COMPLIES WITH NAMEPLATE RATING. 

C.

 

PLUG UNIT INTO GROUNDED ELECTRICAL OUTLET

 

ONLY. 

D.

 

IF  THE  ELECTRICAL  SUPPLY  CORD  IS  DAMAGED,  IT  MUST  BE 

REPLACED BY A CORD WITH AN EQUAL OR HIGHER RATING. 

 
Failure to follow the above precautions could result in electrical shock, fire, personal 

injury, or damage to the unit, and will void the warranty. 
 

 

4.

 

Technical specifications 

  

 
 

Standard Features 

 

™

 

Programmable Digital Temperature Controller 

™

 

Low and High Temperature Alarm Signals 

™

 

Low Fluid Level Indicator  

™

 

Fluid inlet and outlet quick disconnect, 3/8” flow female valves (see 
section 7 for recommended system male connectors) 

 

MRC general series specifications 

Input voltage, VAC 

115-230VAC  

Operating environment temperature range, 

°

C (

°

F)

1

 

4.4 to 45 (40 to 113) 

Control temperature range, distilled H

2

O as coolant,  

°

C (

°

F)

2

 

2 to 40 (35 to 104) 

Control temperature range, using 75% distilled H

2

and 25% glycol or alcohol as coolant, 

°

C (

°

F)

2

 

-12 to 40  

(10.4 to 104) 

Controller accuracy, 

°

± 

0.16

 

Frequency, Hz 

50/60 

1

 

For ambient conditions outside this range, please contact Laird Technologies. 

2

 For temperature settings below 2

°

C, an appropriate coolant must be used.

 

Содержание MRC 150

Страница 1: ...004 Revision 07 MRC 150 300 Thermoelectric Re circulating Liquid Chiller User s Manual USA 1 888 246 9050 Europe 44 1342 315044 Asia 886 3 312 9292 Website www lairdtech com Email clv customerpos lair...

Страница 2: ...the cooled product This manual provides the necessary general information for proper installation and operation and has been developed to achieve the best results when using the Laird Technologies ch...

Страница 3: ...REPLACED BY A CORD WITH AN EQUAL OR HIGHER RATING Failure to follow the above precautions could result in electrical shock fire personal injury or damage to the unit and will void the warranty 4 Techn...

Страница 4: ...19 5 35 1 12 0 7 7 13 8 39 1 20 3 33 8 15 4 8 0 13 3 Weight kg lbs 10 9 24 13 6 30 Maximum flow rate LPM GPM 4 2 9 0 76 3 3 0 87 3 Capacity rating is given at a temperature of 25 C 77 F for the ambie...

Страница 5: ...86 3 312 9292 www lairdtech com clv customerpos lairdtech com MRC 150 Performance Charts Therma T Outlet Fluid Temperat 30 25 20 15 10 5 0 5 10 15 0 50 Capac T C 100 200 300 400 500 600 Capacity BTU h...

Страница 6: ...2 9292 www lairdtech com clv customerpos lairdtech com MRC 300 Performance Charts Thermal Capacity T Outlet Fluid Temperature Ambient Temperature 30 25 20 15 10 5 0 5 10 15 0 50 100 150 200 250 300 35...

Страница 7: ...Doc 60 70 004 Revision 07 Page 7 User s Manual USA 1 888 246 9050 Europe 44 1342 315044 Asia 886 3 312 9292 www lairdtech com clv customerpos lairdtech com...

Страница 8: ...culating in the reservoir otherwise turn off the chiller and wait another 5 minutes See Note Below 9 If there is any doubt that the system has been purged of air then check the fluid level and fill re...

Страница 9: ...acturer is included with each unit 1 To display set point Press and hold the button The display will now show the temperature set point and alternately the units C or F 2 To enter new set point Factor...

Страница 10: ...coolant freezing point High Temp Alarm1 S P 2 A then d U h i C or F 5 C 400 C Customer defined Low Temp Alarm1 S P 2 A then d U l o C or F 1 C 200 C Customer defined Hysteresis1 Alarm Differential b n...

Страница 11: ...roperly 3 Consult Factory Unit is running but the Low Fluid Indicator LED is lit Insufficient fluid level in reservoir Fill up reservoir following air purge instructions Section 5 of this manual Unit...

Страница 12: ...ct to Laird Technologies freight prepaid Laird Technologies in turn will return the warranted product to customer freight prepaid if within the time period specified above Warranty coverage is in effe...

Страница 13: ...s for a particular purpose Laird Technologies will in no way be liable for incidental or consequential damage whatsoever or for any other claims by the customer for damages in connection with Laird Te...

Отзывы: