VitalsBridge Directions for Use (for US Government Customers)
31
rev. B.10
PN 10229
Note that some vital signs monitors filter and average the end tidal CO
2
value, and there may be a
several second delay for the reading to change. Additionally, small CO
2
set values may be difficult to
read on the monitor. A useful feature, available in some patient vital signs monitors, is to "freeze" the
waveform and investigate it more closely (refer to the vital signs monitor documentation for additional
information).
Repeat for all calibration points. Once completed, the calibration data will automatically be
stored/retrieved on the VitalsBridge.