Upright with Newborn PEEP

Product Overview

To dismantle Upright with Newborn PEEP, follow steps 1-6. To reassemble, follow the steps in reverse.

Pull Out

Unscrew

Lift Out

“LAERDAL”

“LAERDAL”

Pull firmly on tab to release

Pull Out

Do not disassemble parts beyond the steps shown. For proper function, make sure to only use parts marked “LAERDAL” (locations as shown above).

Cleaning and Disinfection Procedure

Dismantle

Clean

Disinfect / Sterilize

Dry and Inspect

Assemble and Test

Always dismantle 

Upright with 

Newborn PEEP 

before cleaning. 

1

2

3

4

Wash all parts in a clean tray 

with clean water and mild soap. 

Use a scrub or brush to remove 

any soil. 

Rinse parts in clean water to 

remove all soil and soap. 

Repeat above steps 

until parts are clean. 

Boiling*

Boil all parts in clean

water for 10 min.

Autoclaving

Sterilize by steam

autoclaving

at 136 °C and

2.0 kg/cm

2

 for

10-20 min.

Dry all parts.

 

Visually inspect each part for damage, cleanliness 

and signs of mineral deposits. If the parts become

coated in limescale immerse in equal parts of water

and white vinegar (3-5%) for 10 minutes. 

Do NOT immerse PEEP membrane. 

Rinse in clean water. Repeat if necessary.

PEEP membrane inspection: 

Hold membrane between fingers and squeeze 

softly. Check that slit opens fully, but not beyond 

rip stoppers. 

If in doubt about PEEP membrane function or 

pressure, test as described in Testing PEEP valve,

Page 1. 

Discard any damaged or worn parts.

Reassemble Upright 

with Newborn PEEP.

Test using steps 

shown in Testing 

Before Use section. 

Caution: 

Make sure to attach 

membrane onto the 

patient connector before 

attaching the cap.

Testing before Use

Lip valve function

Pressure release valve and Inlet valve function

Product sealing

PEEP membrane function

Squeeze the bag. Check that the yellow valve 

opens and closes with every squeeze.

Seal the mask with a hand. Squeeze the bag forcefully. 

Check that air is released from the pressure release 

valve. Keep the mask sealed against hand. Release the 

squeezed bag. Check that the bag re-expands without 

resistance.

Keep the mask sealed against the hand. 

Press the pressure release valve down.

Squeeze the bag and check that there 

is no leakage.

Page 2

If any of the above tests fail, dismantle Upright with Newborn PEEP, inspect the components, reassemble and repeat the complete “Testing Before Use” procedure (Section 3).

Remove damaged parts from use.

Reprocessing instructions

Hold Upright with Newborn PEEP firmly at 

the yellow neck label. Press other hand 

upwards against the mask’s sealing cuff so 

that air inside the mask is pushed into 

Upright and out the expiration port with the 

PEEP membrane. Check that the PEEP 

membrane slit opens and seals fully.

Repeat 3 times.

Ensure that the 
silicone bag's lip 

is not twisted or 
tucked under. 

Unscrew

Pull Off

2.b

2.c

Unscrew

Rip Stoppers

OR

 

 

10 minutes

100 °C 

Steam 136 °C

10-20 minutes

*

Validated at 

approximately 

sea-level pressure.

Att 1 to 00021915.pdf   2   12.08.2016   10:09:32