30
31
ADULT PULSES ARE AUTOMATICALLY ADJUSTED TO BP AS FOLLOWS:
Infant pulses are automatically set to absent when Systolic BP is below 10 mm Hg. Otherwise Infant pulses are normal.
ADULT
MANIKIN
CHILD
MANIKIN
INFANT
MANIKIN
Default Blood Pressures (mmHg)
120/80
100/70
94/66
SYSTOLIC BP
CAROTID PULSE
RADIAL/BRACHIAL PULSE
>= 88
Normal
Normal
< 88
Normal
Weak
< 80
Normal
Absent
< 70
Weak
Absent
< 60
Absent
Absent
BLOOD PRESSURE
Pressure range
0 - 300 mmHg
Accuracy
+/- 4 mmHg
Calibration
Pressure sensor must be calibrated to sphygmomanometer. See Preferences menu.
PULSE
Available pulses
Carotid, Brachial, Radial and Umbilical (SimNewB)
Only 3 pulses can be palpated simultaneously. (Power considerations).
Palpated BP
Radial pulse turns off at systolic BP.
Brachial pulse turns off at 20 mmHg, to prevent noise generation in auscultation area.