GUIDELINES ON BEST PRACTICES ON PPH FUNCTIONALITY

               

 

 

Introduction 

During Postpartum Hemorrhage (PPH) simulation scenarios, 
some parts of the simulator inside the pelvis can be damaged 
by exposition to large amounts of fluids. To avoid this, Laerdal 
Medical AS recommends:

• 

The installation of a new power hatch plug, to use during 
every PPH simulation.

• 

Guidelines on how to set up the simulator before starting 
PPH simulation. 

Installing the power hatch plug

This new accessory is intended to stop fluid from  
entering the SimMom pelvis during Postpartum Hemorrhage 
simulation scenarios.

1. Open the power hatch located on the side of the pelvis.

To avoid excessive pooling of fluids in the bed, spread an underpad 
under the simulator’s pelvis.

2. Insert the power hatch plug in the power hatch.

5. Check that the power hatch can close properly with the 
power hatch plug inside.

3. Press to keep it in place.
4. Close the power hatch.

 

   

        Note:  After every PPH simulation, sponge out the blood 
gathered inside the power hatch.

 

   

         Note: Sponge out the excess of blood accumulated in the 
birthing canal before turning the simulator on its side

Setting up the simulator for  

PPH simulation

 

Caution:

• 

Do not open the service panel located on the back side  
of the pelvis. The service panel should only be removed 
and installed by a qualified Laerdal Service Technician.

• 

Do not perform a PPH simulation when the simulator’s 
head is down. Make sure the head and upper body of the 
simulator are raised during PPH simulation to ensure the 
correct flow of blood through the simulator. 

Содержание Limbs & Things SimMom

Страница 1: ...SimMom Guidelines on Best Practices on PPH Functionality EN NL FR PL IT ZH ES JA BR KO...

Страница 2: ...s in the bed spread an underpad under the simulator s pelvis 2 Insert the power hatch plug in the power hatch 5 Check that the power hatch can close properly with the power hatch plug inside 3 Press t...

Страница 3: ...s dans le lit installez une al se sous le bassin du simulateur 2 Ins rez l obturateur de trappe d alimentation dans la trappe d alimentation 5 V rifiez que la trappe d alimentation peut se fermer corr...

Страница 4: ...mpapador debajo de la pelvis del simulador 2 Inserte el nuevo tap n de la ventana de alimentaci n en la ventana de alimentaci n 5 Compruebe que la ventana de alimentaci n puede cerrarse correctamente...

Страница 5: ...ulo di fluidi nel letto mettere una traversa sotto il bacino del simulatore 2 Inserire il coperchio nel vano dell alimentazione 5 Controllare se il portello dell alimentazione si chiude correttamente...

Страница 6: ...uidos na cama estenda um forro sob a pelve do simulador 2 Insira o tamp o no compartimento de energia 5 Verifique se o compartimento de energia pode ser fechado corretamente com o tamp o dentro 3 Pres...

Страница 7: ...een onderlegger onder het bekken van de simulator om overmatige vochtophoping in het bed te voorkomen 2 Steek de stekker van het voedingsluik in het voedingsluik 5 Controleer of het voedingsluik goed...

Страница 8: ...boku miednicy Aby unikn nadmiernego gromadzenia si p yn w w ku pod miednic symulatora nale y roz o y podk ad 2 W o y wtyczk komory zasilania do komory 5 Sprawdzi czy komora zasilania daje si poprawnie...

Страница 9: ...PPH PPH Laerdal Medical AS PPH PPH PPH SimMom 1 2 5 3 4 PPH PPH Laerdal PPH PPH...

Страница 10: ...PPH PPH Laerdal Medical AS PPH PPH SimMom 1 2 5 3 4 PPH PPH Laerdal PPH PPH...

Страница 11: ...PPH PPH Laerdal Medical AS PPH PPH SimMom 1 2 5 3 4 PPH PPH Laerdal PPH PPH...

Страница 12: ...20 20856 Rev B 2023 Laerdal Medical AS All rights reserved...

Отзывы: