11
ENGLISH
2. To release the Canister Holder, press down
the Release Arm while sliding the Holder
towards you. Remove the Canister from the
Holder.
3. Release the Patient Suction Tubing
(if applicable).
4. Safely dispose of the contents of the Canister
according to local protocols.
5. Clean the LSU according to section 6.3 and its
reusable parts according to section 6.4.
4.3 Assembly of the Reusable Canister System
1. Place the yellow Gasket in the Lid, make sure
that the gasket is placed flat in the Lid and
not twisted.
2. If the Float Ball has been removed, hold the
lid upside down and snap the ball into place
as shown.
Note: Never use the LSU without the Float Ball.
3. Secure the Lid to the Canister by turning it
clockwise until it is securely fastened.
4. Connect the Suction Catheter Adapter to
the Patient Tube.
5. Connect the Angled Connector to the Patient
Tube.
6. Mount the yellow Vacuum Seal in the centre
hole of the Lid and push it into place.
7. Connect the Aerosol Filter (blue ring) to
the blue Vacuum Inlet on the LSU.
(Color code: Blue on Blue)