6 OPERATION
LACO Technologies
TITAN VERSA Operations and Maintenance Manual
51
11.
Verify the leak rate is stable. Calibration verifies background reading.
12.
If the unit fails a calibration the user will be notified. See Calibration History for
more information on reviewing calibration results.
Vacuum Method Leak Testing
6.4.2.1.
Outside-In Leak Testing
In this method, the test object is evacuated (direct connection to vacuum inlet test port
KF-25) and the user sprays helium to verify test leak integrity. This method allows for
leak location.
If test objects are small enough, a plastic bag can be placed over the test object and
helium sprayed into bag. This method ensures helium application is global around the
entire part
Содержание TITAN VERSA C
Страница 1: ...TITAN VERSA LEAK DETECTORS OPERATIONS AND MAINTENANCE MANUAL ...
Страница 10: ......
Страница 28: ...4 DESCRIPTION 18 TITAN VERSA Operations and Maintenance Manual LACO Technologies Vacuum Schematic Main Version ...
Страница 77: ...6 OPERATION LACO Technologies TITAN VERSA Operations and Maintenance Manual 67 Method Settings ...
Страница 82: ...6 OPERATION 72 TITAN VERSA Operations and Maintenance Manual LACO Technologies AutoTest Settings ...