background image

   

  - 1 - 

 

Operating Manual Number : LBI-500EE 

Edited : December 1999 

Revised : January 2016 

July 2017 

May 2018 

 
 
 
 
 
 

  Low Temperature BOD Incubator 

Model LBI-150E / LBI-250E / LBI-500E / LBI-1000E 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

Содержание LBI-1000E

Страница 1: ...1 Operating Manual Number LBI 500EE Edited December 1999 Revised January 2016 July 2017 May 2018 Low Temperature BOD Incubator Model LBI 150E LBI 250E LBI 500E LBI 1000E...

Страница 2: ...described in this manual and operation procedures are not taken properly it may result in malfunction of the equipment Note This mark is to indicate precautions which must be taken properly for safe...

Страница 3: ...match it may cause fire or electric shock and if used in combination with other equipment it may cause abnormal heat generation and fire The figures diagrams and numeric values used in this manual are...

Страница 4: ...of contents 1 Tables of contents 4 2 Product overview 5 3 Product specifications 6 4 Part and functions 7 5 Installation 9 6 Operating 10 7 Maintenance and repair of products 12 8 Cause and trouble s...

Страница 5: ...bility ensure user safety This equipment equipped with a compressor which allows general incubation as well as low temperature incubation experiments This equipment is made with high quality stainless...

Страница 6: ...0 1 Uniformity 0 5 at 20 Controller Microprocessor PID Multi Function Controller Display 4 Digits LED Display Timer 99hr 59 min Continuous Material Inner Stainless Steel SUS304 Outer Epoxy Powder coat...

Страница 7: ...is in operation Heater operation indicating lamp OUTPUT Illuminates the heater is in operation Timer display lamp TIMER Illuminates when timer is operating START STOP button START STOP Button to start...

Страница 8: ...heater output off when there is over temperature due to equipment malfunction Set this 10 20 higher than the set temperature Over temperature protection device is analog type therefore it will not acc...

Страница 9: ...lease install AVR Automatic Voltage Regulator separately For electric safety the unit should be earthed If the power socket is fitted with earth cord please check the outlet for earth before use Do no...

Страница 10: ...erature will be indicated on the display waiting for input 4 Press LEFT SHIFT button to move to digit you want change value 5 Press INC button to change value 6 4 How to set time 1 Press MODE button a...

Страница 11: ...ication lamp illuminates and the timer starts to count down 3 After set time Timer indicating lamp turns off and operation stops 6 7 Auto tuning 1 Auto tuning function automatically sets optimal P ID...

Страница 12: ...the compressor If the sound of equipment operating suddenly becomes loud or irregular please check with our service team 7 2 Maintenance of electric compartment Be careful not to let water enter the e...

Страница 13: ...eaker short circuit check Switch on the circuit breaker When temperature control is not possible Check the condition of the controller Calibrating regulator settings Over temperature protection check...

Страница 14: ...14 APPENDIX Consumables List NO PART CODE NAME SPECIFICATION UNIT Q ty 1 LBI E001 HEATER LBI 150E M 750W EA 1 LBI 250E M 1KW EA 1 LBI 500E M 750W EA 2 LBI 1000E M 750W EA 3 2 3 4 5 6 7 8 9 10...