AQ-00275-000, Rev. 3
13
5. Save the scan to a special directory on your computer.
Reference Beam Energy Scan:
1. Set up instrument parameters as listed on the Labsphere diagnostic scan reports in the
quality assurance documentation.
2. Change the instrument configuration to
Single Front
beam mode from the Setup Dia-
log Box.
3. Place a Spectralon standard over the reflectance sample and reference ports.
4. Initiate the scan.
5. Save the scan to a special directory on your computer.