The features under [MENU] will allow you access useful information, and also allow
you to change many settings.
Functions Description
CHART
Last 24 hours temp recording in a line format.
LOG
Numbers of data recorded.
MIN/MA
X
1-point maximum and minimum temp record. Press [RESET] to clear
the history.
SUPPORT
Tech Support contact information and QR code link to download
manuals.
TEST
Press [TEST] to set off visual, audible, and remote alarm contact to
validate the alarms are working.
SETTING
Enter password (factory: 0,0,0,0) to change the following settings:
°C/°F
Change the display scale between °C or °F.
HI/LO
Change the high & low temp alarm point.
OFFSET
Change the display readout offset.
TIME
Change the display time zone.
VOLUME
Change the audible alarm volume.
RINGBACK DELAY
Change the audible alarm reminder.
REMOTE DELAY
Change the remote alarm delay.
DOOR DELAY
Change the door alarm delay.
USERS
Change the password code.
CHECK 1-POINT TEMPERATURE HISTORY
Your facility policy may require to record this refrigerator’s maximum and minimum
temperature point within last 24 hours. Press [MENU] -> [MIN/MAX] to check the
1-point temp history record. Press [RESET] to clear the history.
TECH SUPPORT INFORMATION
A QR code will allow you to locate Technical Support Info. In addition, the display
shows our technical service department contact phone number.