6.6 Infusion Volume Transducer
The infusion volume transducer is a part of the NXT Go Infusion Hook. The transducer allows
Synergy to measure the fluid used to infuse the bladder during a Urodynamic study. The
infusion volume transducer (Figure 80) is setup as follows:
Hang the NXT Go Infusion Hook from an IV Pole. Ensure the NXT Go Infusion Hook
connector cable is attached to the NXT Go.
Hang a 1000 ml bag of sterile saline or water on the NXT Infusion Hook.
Connect the Infusion Pump Tubing to the saline bag.
IMPORTANT
:
Ensure that the infusion line is not pulled tight or the transducer reading will not be
accurate.
Do not allow the saline bag to touch any other mechanical component or module of the
system. The saline bag must free hang from the NXT Go Infusion Hook to allow the
transducer to measure an accurate weight.
Figure 80: Infusion Volume Transducer on IV Pole
The infusion transducer is calibrated for up to 1.5kg of fluid. Ensure the
Fluid Density
setting captured in the workflow aligns with the fluid type being used. Fluid density must be
set in the workflow configuration prior to beginning the study. Refer to the
6.7 Roam NXT
The Roam NXT is a wireless capable module intended to be connected to the detachable PIM
NXT models. Together, the devices are used to record pressure and EMG during
Urodynamic study. The PIM NXT must be connected to the Roam NXT to facilitate
functionality of both devices. The devices may be connected directly or by using the Roam
NXT to PIM NXT Connector Cable.
PIM NXT
The PIM NXT is the connection point for patient pressure and EMG inputs. Ensure to select
the PIM NXT model that facilitates use of the pressure catheters deployed, PIM NXT for T-
DOC or PIM NXT for Fluid.