Safety Information
Nexam Pro User Manual
247
CAUTION
Change the battery only when the patient is
disconnected from the WPM (unplug catheter and all
patient leads from the WPM).
CAUTION
After low battery indication, exchange the batteries within
15 minutes.
CAUTION
Remove the batteries when the wireless patient module
is not used for a longer period (month).
NOTE
Environmental conditions (as thick walls/doors, iron in
walls/doors) may compromise the Bluetooth connection
of the wireless patient module and battery lifetime.
Wireless patient unit (WPU-L4
/ WPU-eS)
The wireless patient unit complies with Part 15 of the FCC Rules. Operation is
subject to the following two conditions: (1) this device may not cause harmful
interference, and (2) this device must accept any interference received,
including interference that may cause undesired operation. This device
contains FCC ID P4IBTM101.
CAUTION
Change the battery only when the patient is
disconnected from the WPU (unplug catheter and all
patient leads from the WPU).
CAUTION
After low battery indication, exchange the batteries within
15 minutes.
CAUTION
Remove the batteries when the WPU is not used for a
longer period (month).
NOTE
The quality of the wireless connection may be
compromised by environmental conditions like human
bodily mass, metal plates, concrete wall reinforcement,
and other wireless connections like Bluetooth, WiFi,
mobile telephones, and so on.