48
Nexam Pro User Manual
Chapter 2
•
LABORIE EMG-PE (= disposable waterproof protected EMG surface
electrodes including lead wires) to be used with EMGC-3DIN-TP-40
(= guarded EMG cable with three x DIN42802 plug and three x DIN 42802
receptacle).
•
EMGC-3DIN-CN200 (short EMG adaptor cable with three x DIN42802 to
female DIN 5 and 200 cm reference lead), to be used for needle EMG
electrodes.
The color coding is as follows:
Red
Signal -
Black
Reference
Green / Yellow
The recommended EMG cable to use with the WPM or Luna is the UD 597 A6
(= guarded EMG cable for three surface electrodes with press studs, 60 cm).
Figure 2.36
EMG cable (for Luna or WPM)
The color coding is as follows:
Red
Signal -
Red
Reference
Black
2.5.11 Prepare the Flowmeter
The flowmeter can be used to measure the voided volume and the flow during
urodynamic investigations. The flowmeter is supplied with wireless Bluetooth
technology, so cabling is not required.
Содержание Nexam Pro CIM
Страница 1: ...Nexam Pro User Manual...