Analysis Program
Nexam Pro User Manual
233
study), because of the adverse effect of the transurethral catheter in women
undergoing pressure-flow studies. The first time the nomogram is displayed,
you can will be prompted to enter the free Qmax obtained with free-flow(s).
The Pdet.max will be obtained from the current study as the maximum
detrusor pressure that was measured during the current study.
The Nomogram is divided into four areas, indicating no obstruction or mild,
moderate, or severe obstruction.
Read also:
Jerry G. Blaivas and Asnat Groutz: Bladder Outlet Obstruction Nomogram for
Women With Lower Urinary Tract Symptomatology,
Neurology and
Urodynamics
19:533-564 (2000).
13.5.12 Composite Nomogram
The composite nomogram shows an x-y plot of the Pdet.Qmax (detrusor
pressure at Qmax) related to Qmax.
Click the
Composite nomogram
button or choose
Options
> Composite
nomogram
from the menu to display the plot.
Figure 13.17
Composite Detrusor/flow plot
The results of the composite nomogram are validated only for male patients.