TECHNICAL SPECIFICATIONS
FILTER SPECIFICATIONS
Type
: Transmitted Fluorescent illumination based on switching of dichroic mirror
cubes
Observation modes
: B-excitation and G-excitation
Lamp type
: HBO 50 W, high pressure Mercury vapor lamp
Supply Voltage
: 110V AC, 60 Hz / 220V AC, 50Hz
Operating environment
: Indoor use
0
0
Ambient temperature 5 C to 40 C
0
Maximum relative humidity 80% at 40 C
B-excitation
Excitation filter
:
Transmission 90% at wavelength 420 to 480 nm.
Transmission 0% at wavelength 485 nm onwards.
Dichroic mirror
:
Transmission 93% at wavelength 540 nm onwards
.
Transmission 0% at wavelength 400 to 530 nm.
Barrier filter
:
Transmission 95% at wavelength 535 nm onwards
Transmission 0% at wavelength 400 to 500 nm.
Recommended Fluorescent Dye
:
Acridine Orange, both DNA & RNA
G-excitation
Excitation filter
:
Transmission 85% at wavelength 490 to 525 nm.
Transmission 0% at wavelength 530 nm onwards.
Dichroic mirror
:
Transmission 90% at wavelength 600 nm onwards
.
Transmission 0% at wavelength 400 to 595 nm.
Barrier filter
:
Transmission 92% at wavelength 615 nm onwards
Transmission 0% at wavelength 400 to 575 nm.
TM
Recommended Fluorescent Dye
:
Acridine Red FM 5-95
Fluorescence Attachment User Guide
9