 SL 20i 

INSTRUCTIONS FOR USE

12

Lock

 

Projected
       Slit 

Figure-12 Focus on Slit

Figure-13 Adjust Patient Height

Base 

Figure-14 Adjust Height

Figure-15 Adjust Slit Width

Canthus 
Mark 

8126700-795 

 SL 20i 

Issue 1.0

Printed on June,2019

OPERATION

1. Turn on the power using the On/Off switch located on the front
     of the power supply. Brightness can be adjusted by rotating the
     illumination level knob. Change the knob for brighten control.

NOTE:  

The maximum position is for intermittent use only.

              Continuous use will shorten lamp life.

2. Insert the Focusing Rod in the pivot post of the instrument
     body to make rough IPD and focus adjustments.

3. Position the light onto the flat surface of the focusing rod and 
   adjust the pupillary distance and focus of the eyepieces to 
     suit the needs of the operator. Refer to figure-12.

4. Set Magnichanger knob to 16x. Bring 10mm Aperture into the 
     light path by rotating aperture disc. Focus image on target rod.
   Circular Image should be in the Centre of FOV as Shown in
     Figure 12.     
                 
5. Using the Slit Width Knobs, adjust the projected slit so that
   the thinnest slit is shown on the Focusing Rod. Refer to fig.
     -12 & 15.        
    

NOTE:  

The thinnest line will allow for greater accuracy.

6. Remove the Focusing Rod.

7. To position a patient, adjust the chin rest height by turning
   the Chin rest Elevation Handle on the post of the Chin Rest 
     Assembly until the patient’s canthus is in line with the canthus 
     mark on the chin rest post. Refer to figure-13.

8. Microscope elevation is adjusted by rotating the joystick and
     observing the slit image through the Microscope Assembly until
     slit is centered on the patient’s cornea. Refer to figure-14.

9. Move the slit lamp with the joystick held firmly and slightly 
     angled toward the patient, until the slit appears sharply on the 
     cornea.

NOTE:  

The accuracy of this rough adjustment should be checked

by the naked eye. The fine adjustment is performed while 
observing the slit through the microscope.

10. Tilt the Joystick, which is now held lightly at its upper end,
     until the slit appears sharply at the depth of the eye which is to     
     be observed. 

11. The horizontal motion of the base can be locked by tightening
      the Base Locking Screw. Refer to figure-14.

Chin rest 

Elevation 

Handle 

Height 

Adjust Lamp 

Slit 

Width/Rotation 

Knob 

Содержание SL 20i

Страница 1: ...Manual Part No 8126700 795 Issue 1 0 Printed on June 2019 To ensure proper use of this instrument as well as to avoid injury while operating Instrument understanding this manual completely before use...

Страница 2: ...Reserves the right to make change in the product described in this manual without notice and without incorporating those changes in any products already sold ISO 13485 Certified LABOMED products are d...

Страница 3: ...ION OF INPUT POWER 11 INSTRUCTIONS FOR USE 12 OPERATION 12 SLIT LENGTH 13 FILTER DIAL 13 SLIT ROTATION 13 CLEANING MAINTENANCE 14 EXTERNAL CLEANING 14 FOREHEAD CHIN REST PREPARATION 14 CLEANING THE GL...

Страница 4: ...O ENSURE CONTINUED SAFE USE OF THIS INSTRUMENT WARNING TO AVOID RISK OF ELECTRIC SHOCK THIS EQUIPMENT MUST ONLY BE CONNECTED TO A SUPPLY MAIN WITH PROTECTIVE EARTH OTHERWISE FAILURE OF INSTRUMENT AND...

Страница 5: ...AY OCCUR SEE THE CARE AND MAINTENANCE SECTION FOR DETAILED CLEANING INSTRUCTIONS CAUTION MEDICAL ELECTRONIC EQUIPMENT NEEDS SPECIAL PRECAUTIONS WITH RESPECT TO ELECTROMAGNETIC CHARGE EMC AND NEEDS TO...

Страница 6: ...Power Protective Earth Connection ON OFF Manufacture Catalog Number Serial Number Waste of Electrical and Electronic Equipment Compliance to medical Device Directive 93 42 EEC Accompanying Document m...

Страница 7: ...ticians and eye care technicians should operate this instrument Please retain this manual for future reference and to share with other users Additional copies can be obtained from your authorized LABO...

Страница 8: ...k Knob 18 Instrument base Lock Knob 19 Guide Rail Cover 20 Geared Rollers 21 Guide Rails 22 Fore head support 23 Power Supply Assembly 24 Table Top 25 Patient Handles 26 Observation Head 27 Chin Suppo...

Страница 9: ...g Item Packed as Shown in Figure 2 a Chin Rest Assembly b Microscope Head with Eyepiece Mounted c Power Supply Box d Base Assembly e Patient Resting Handle f Illumination Arm Assembly g Accessories Sl...

Страница 10: ...ck the base at any position to prevent any undesired movement by using Base Locking Knob Shown as A in Fig 5 8 Assemble Illumination moving arm assembly onto the base using Allen Screw M6 as shown in...

Страница 11: ...MINER OR A DANGER TO THE PATIENT WARNING POSITION THIS INSTRUMENT SO THAT IT IS NOT DIFFICULT TO OPERATE THE DISCONNECTION DEVICE PLUG 1 After the unit is in its secure location apply the correct inpu...

Страница 12: ...bs adjust the projected slit so that the thinnest slit is shown on the Focusing Rod Refer to fig 12 15 NOTE The thinnest line will allow for greater accuracy 6 Remove the Focusing Rod 7 To position a...

Страница 13: ...dicated in the scale of the slit lamp arm Refer to figure 17 15 Magnification is altered by using lever on the Magnichanger Refer to figure 16 SLIT LENGTH The Slit length is adjusted by rotating the S...

Страница 14: ...dampened with a mild soap and water solution CHANGING THE HALOGEN BULB WARNING NEVER REMOVE A BULB THAT HAS RECENTLY BEEN IN USE AS IT WILL BE VERY HOT WAIT UNTIL IT HAS COOLED AND USE GLOVES OR A TH...

Страница 15: ...y FUSE REPLACEMENT Replace the fuses in the Power Input Module with the fuses indicated in the Specifications section of this manual 1 Remove input power to the instrument 2 Press down on the top tab...

Страница 16: ...bulb housing Double slit visible in microscope Microscope not focused on focusing rod before use Install focusing rod and check to ensure microscope is focused on it Bulb not installed properly Check...

Страница 17: ...80 Atmospheric Pressure 80 kPa 23 6 in Hg to 106 kPa 31 3 in Hg Transportation Storage 10 Temperature 20 C 4 F to 70 C 158 F 106 kPa Relative Humidity 10 to 80 non condensing Atmospheric Pressure 50k...

Страница 18: ...nnecting the equipment to a protective grounding system so that the accessi ble metal parts do not become conductive in case of failure in the basic insulation Grade of protection against a hazardous...

Страница 19: ...T H S ADAPTER KIT SBS WITH VIDEO OUTPUT CCD ADAPTER 8122500 8122600 800 6134131 6134110 6134120 6134172 I PHONE 6 6134172 1 6 I PHONE 6 PLUS 6134172 1 6P 8122600 851 8122600 852 CCD CAMERA I PHONE 7 6...

Страница 20: ...Test Compliance Electromagnetic Environment Guidance RF Emissions CISPR 11 Group 1 Class A Harmonics IEC61000 3 2 Class A Flicker IEC61000 3 3 Complies 8126700 795 SL 20i Issue 1 0 Printed onJune 201...

Страница 21: ...5 Cycle 60 Dip for 5 Cycles 30 Dip for 25 Cycles 95 Dip for 5 Seconds 95 Dip for 0 5 Cycle 60 Dip for 5 Cycles 30 Dip for 25 Cycles 95 Dip for 5 Seconds Power Frequency 50 60Hz Magnetic Field IEC61000...

Страница 22: ...ccording to the transmitter manufacturer and dis therecom mended separation distance in meters m Field strengths from fixed Rf transmitters as determined by an electromagnetic site survey should be le...

Страница 23: ...or use in the electromagnetic environment in which radiated RF disturbance are controlled The customer or user of the SL 20i can help prevent Electromagnetic interference by maintaining a minimum dist...

Страница 24: ...t Labomed shall not be liable for any special Incidental or consequent damages for any negligence breach of warranty strict liability or any other damages resulting from or relating to design manufact...

Страница 25: ...Revision History Rev No Date of Release Change App By DCR...

Страница 26: ...4538 U S A Phone 510 445 1257 Fax 510 991 9862 E mail sales laboamerica com www laboamerica com Labomed Europe Essebaan 50 NL 2908 LK Capelle a d IJssel The Netherlands Tel 31 0 10 4584222 Fax 31 0 10...

Отзывы: