BA_054 Rev 1K7 Resistance thermometer 

Page 3/3 

 

3.1  Electrical Connection 

Complete the mechanical installation before you connect the device electrically. 

Set up all electrical connections while the voltage supply is switched off. 

  

 

 

 

M12 circular connector 

cable gland 

3-wire circuitry 

transmitter 

3-wire circuitry 

ϑ

 

(+)

(-)

n.c.

n.c.

 

2

6

ϑ

1

 

 

Figure 1:  Variants of electrical connection 

  

Route the cable so that it does not apply force or torque to the measuring insert. 

For  devices  with  field  housing,  hold  the  housing  with  the  other  hand  when  you  screw  or 
unscrew the field housing cap. 

  

Operation 

During operation, take care that the device remains within its intended pressure and tem-
perature ranges. No other monitoring is necessary. 

  

Permissible media temperature:  

-40…150 °C 

  

Permissible ambient temperature:  

-40…100 °C 

When supplied with a transmitter, the permissible ambient temperature is reduced to 85°C. 

  

Restricted  ambient  conditions  may  apply  in  explosion  protected  environments  (see 
XA_001). 

  

4.1  Maintenance / Service 

When  properly  installed  in  accordance  with  applicable  specifications,  this  device  is 
maintenance-free. However, we recommend an annual recalibration of the device. 

  

Disassembly 

When measuring hot media, make sure that the device has cooled down prior to any dis-
mounting or wear appropriate protective clothing to avoid burns. 

  

Switch off the power supply to the device before disconnecting the electrical connections. 
Once this is done, the device may be mechanically removed. 

  