MultiGene Thermal Cycler
19
The temperature and time ranges are shown below.
Temperature range: 4.0 to 99.9
◦
◦
C
Time range (minute: second): 00:01 to 99:59 ( 99:99 =
∞
)
6.2 Inserting/Deleting Temperature Segment
Temperature Segment can be inserted or deleted by selecting
F2:INSERT
or
F3:DELETE
key. The inserted or deleted temperature segment is dependent upon the
location on the temperature segment where the cursor is positioned. If the cursor is
positioned inside the first multi-temperature/time segment, pressing
F2/F3
will
insert/delete the whole Multi-Temperature Cycle. If the cursor is located in any other
Multi-Temperature/Time segment only the selected segment will be inserted/deleted.
6.3 Changing Temperature Segment Count (Temp)
and Cycle Count (Cyc)
The Temperature Segment Count(Temp) can be modified by entering a new value (1 to 9)
in the first Temperature position and pressing ENTER. When the Temperature Segment
Count is larger than one, the temperature Cycle Count (Cyc) will appear. The
Multi-Temperature Segment will cycle during this segment. The maximum temperature
segment per protocol is 12 segments. The temperature segment count and cycle count
ranges are shown below.
Temperature Segment Count(Temp.) range: 1 to 9
Cycle Count(Cyc) range: 1 to 99
6.4 Advanced Settings
Users can set up the following advanced functions: 1)
Gradient
, 2)
Temperature/Time
Increment/Decrement
, 3)
Pause
,.
F4:ADVANCE
button, is used to enter these settings. The screen below will be
displayed.