Labnet Spectrafuge 16M Скачать руководство пользователя страница 6

 

 

 
Adapter for 0.5ml tubes 

Order 

no.  

 

 

C1205 

Tube measurement   

 

8 x 30mm 

Max. speed   

 

 

14,000rpm 

Centrifuging radius   

 

6.6cm 

RCF (g-value)  

 

 

14,462 x g 

 
Adapter for 0.4ml tubes 

Order no.   

 

 

C1206 

Tube measurement   

 

6 x 47mm 

Max. speed   

 

 

14,000rpm 

Centrifuging radius  

 

7.3cm 

RCF (g-value)  

 

 

16,000 x g 

 
Adapter for 0.2ml tubes 

Order no.   

 

 

C1222 

Tube measurement   

 

6 x 21mm 

Max. speed   

 

 

14,000rpm 

Centrifuging radius   

 

6.1cm 

RCF (g-value)  

 

 

13,366 x g 

 

Rotor maintenance 

The rotor should be cleaned thoroughly after each use. Thorough cleaning must 
be performed when spinning samples containing phenol or phenol chloroform. 
Periodically inspect the rotor for dents, dings, scratches, discoloration and cracks. 
If any damage to the rotor is found, discontinue use of the rotor immediately and 
replace. 
 

Mounting and securing the angle rotor 

Remove the rotor screw from the motor shaft by turning the screw 
counterclockwise. Clean the motor shaft and the rotor mounting hole (see figures 
1&2). Place the rotor on the motor shaft ensuring that the cross-pin (figure 1) 
aligns correctly with the rotor slot (see figure 2). Note: Figures 1 and 2 are 
located on the following page. Reinstall the rotor screw on the motor shaft by 
turning it clockwise. Hold the rotor with one hand and hand-tighten the rotor 
screw. Use an adjustable or 1/4 inch wrench (some units are supplied with a 
wrench) to tighten the screw When loading the rotor, refer to figure 3 (located on 
page 5). Loading in the pattern indicated will ensure a balanced load. Tubes to 
be loaded should be filled equally by eye. The difference in the weight between 
the tubes should not exceed 0.1 gram. A partially loaded rotor may be 
centrifuged if the loading scheme for balancing a rotor given in figure 3 is 
followed. 
 

 
 
 
 

Содержание Spectrafuge 16M

Страница 1: ...Lit Code M00001 Rev 2 August 2016 Spectrafuge 16M Microcentrifuge User Manual C0160 C0160 100V C0160 230V...

Страница 2: ...illage may harm unit NEVER put hands in the rotor area unless the rotor is completely stopped NEVER move the centrifuge while the rotor is spinning NEVER use solvents or flammables near this or other...

Страница 3: ...tors and accessories Rotor maintenance Mounting and securing the angle rotors Removing the rotor Overloading rotors Operation 5 Closing the lid Lid release Lid lock Speed selection Selection of operat...

Страница 4: ...mpaired Do not operate the centrifuge if any of the following conditions exist The centrifuge has not been installed properly The centrifuge is partially dismantled Service has been attempted by unaut...

Страница 5: ...table laboratory bench cabinet etc To guarantee sufficient ventilation ensure that the centrifuge has at least 15cm 6 inches of free space on all sides including the rear The centrifuge should not be...

Страница 6: ...ting and securing the angle rotor Remove the rotor screw from the motor shaft by turning the screw counterclockwise Clean the motor shaft and the rotor mounting hole see figures 1 2 Place the rotor on...

Страница 7: ...4 Figure 1 Chamber and motor shaft Figure 2 Bottom of angle rotor Figure 3 Loading the rotor...

Страница 8: ...igher density it is necessary to reduce the speed Failure to reduce the speed may result in damage to the rotor and centrifuge The revised maximum speed can be calculated with the following formula Re...

Страница 9: ...labeled speed The scale is directly proportional to the speed a setting of 9 corresponds to 9 000rpm a setting of 13 corresponds to a speed of 13 000rpm etc Selection of operating time and momentary...

Страница 10: ...The rotor should be cleaned after each use When spinning 8 samples containing phenol or phenol chloroform the rotor should be cleaned immediately after use Disinfection Should a spill of infectious ma...

Страница 11: ...lose lid correctly Possible reason No speed or time has been selected Solution Set speed and or time Where to call Should you have any questions about the Spectrafuge 16M or its accessories please cal...

Страница 12: ...l Force chart can be used to determine g values To use this chart find the radius value on the radius scale Place the edge of a ruler on the value Place the right side edge of the ruler on the speed s...

Страница 13: ...s symbol identifies a Protective Earth PE terminal which is provided for connection of the supply system s protective earth green or green yellow conductor EQUIPMENT DISPOSAL EUROPEAN REGULATIONS Acco...

Страница 14: ...paint or finish Claims for transit damage should be filed with the transportation carrier ALL WARRANTIES INCLUDING THE IMPLIED WARRANTY OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE ARE LIMI...

Страница 15: ...NOTES...

Страница 16: ...9299560000 31 Mayfield Ave Edison NJ 08837 P 732 417 0700 F 732 417 1750 Toll Free 888 522 6381 labnetinfo corning com www labnetinternational com...

Отзывы: