222DS Benchtop Shaking Incubator
User Manual
I-5222-DS
I-5222-D-230V
Страница 1: ...222DS Benchtop Shaking Incubator User Manual I 5222 DS I 5222 D 230V ...
Страница 2: ...ERAL SAFETY RECOMMENDATIONS 5 4 0 INSTRUCTION FOR USE 6 4 1 INTRODUCTION 6 4 2 BASIC OPERATIONS 7 4 3 CHANGING PARAMETERS DURING OPERATION 9 4 4 SHAKING WITHOUT HEATING 10 4 5 HEATING WITHOUT SHAKING 10 5 0 TROUBLESHOOTING 10 5 1 ERRORS 10 5 2 DESCRIPTION OF POSSIBLE ERRORS 11 6 0 TECHNICAL SPECIFICATION 11 6 1 TECHNICAL DATA 11 7 0 TEMPERATURE RE ADJUSTMENT 12 8 0 MAINTENANCE AND CLEANING INSTRUC...
Страница 3: ...ast parameter values reside in the memory even if the device was switched off Precisely adjusted and maintained temperature regulation within 0 5 C of absolute deviation from the set temperature in device working area Incorporated HOLD and STEP time function with 1 minute resolution Special parallel temperature protection for your safety Error number combined indication of possible errors Easy int...
Страница 4: ...ed during transportation immediately make a complaint to the carrier Any incorrect delivery or missing parts should be reported to the distributor 3 2 SELECTING THE RIGHT PLACE When selecting the right place for device please consider the following Put the device on a smooth horizontal and stable place Leave enough space on the back side of the device for normal air circulation Leave enough space ...
Страница 5: ...ers and set new parameters values if necessary Please refer to procedures described in next chapters 3 6 GENERAL SAFETY RECOMMENDATIONS Before any installation read this User s Manual Before cleaning disconnect the device from mains power supply Device could be cleaned with a soft lightly wet cloth Don t use cleaners or liquids on aerosol basis Do not use the device nearby water sources Take care ...
Страница 6: ...al light shines when the shaking incubator is set on time 6 ENCODER with rotating encoder clockwise or counter clockwise you can change Time Speed and Temperature on the shaking incubator Push encoder to change between Time Speed and Temperature If you rotate encoder knob fast the value on display goes up or down quickly 7 START STOP key START or STOP operating 8 TEMPERATURE display 9 SPEED TIME d...
Страница 7: ...set temperature to desired value 37 5 37 5 C POWER If you want to use the shaking incubator without temperature regulation set OFF on display You get OFF function on display when you rotate encoder under 0 0 C or above 70 0 C POWER Press encoder knob Speed signal light blinks With rotating encoder clockwise or counter clockwise set rotating speed to desired value 34 34 RPM 120 120 RPM POWER ...
Страница 8: ...nction on display when you rotate encoder under 0 01 min or above 99 0 h Press START STOP key once to exit the changing of parameters You are now in starting mode Only signal light of the last changed parameter shines Note You always change the parameter whose signal light blinks temperature speed or time You can always change one two or three parameters To exit the changing of parameters always p...
Страница 9: ...rking parameters When time elapses or you press START STOP key again message End appears on display and Run signal light pulses When shaking incubator completely stops it places itself to last used values for Time Speed and Temperature If you lift the lid during operation shaking incubator stops immediately and message Opn Lid appears on displays When you close the lid again shaking incubator star...
Страница 10: ...ture sensor notices 5 C higher temperature than set temperature Shaking incubator is automatically stopped Call nearest service E 23 Set temperature is not reached in 2 hours Shaking incubator is automatically stopped Call nearest service 6 0 TECHNICAL SPECIFICATIONS 6 1 TECHNICAL DATA Heater power 700 W Fan power 17 W Shaking motor power 35 W Fuse 2 x 3 15A 250V 2 x 6 3A 115V Environment temperat...
Страница 11: ...onds message Cor correction appears on display Press encoder knob once to enter the value setting procedure and set the value with rotating encoder counter clockwise or clockwise see Example 1 or Example 2 Pay attention to the sign of the value plus or minus When you set the value press encoder knob again Press START STOP key to finish the temperature re adjustment Note Temperature re adjustment p...
Страница 12: ...nner gasoline and similar Avoid any contact with these chemicals Do not use detergents with high levels of alcohol It is the best to use 2 aqueous solution of non aggressive cleaning products dishwashing detergent Wipe the surface with a lightly damp soft cloth Remove dust and dirt with dry soft cloth Take care when cleaning the device after operating especially when you have used the device at te...
Страница 13: ... equipment at lifecycle end by handing it over to an authorized facility for separate collection and recycling It s also you responsibility to decontaminate the equipment in case of biological chemical and or radiological contamination so as to protect from health hazards the persons involved in the disposal and recycling of the equipment For more information about where you can drop off your wast...
Страница 14: ...declaration becomes invalid Device Name Benchtop Shaking Incubator 222DS Device Models Numbers I 5222 DS I 5222 DS 230V Relevant EC Directives Low Voltage Directive 2006 95 EC Machinery Directive 2006 42 EC EMC Directive 2004 108 EC WEEE 2002 96 EC Harmonized Standards EN ISO 12100 2 EN 61010 1 EN 61010 2 010 EN 61010 2 051 EN 55011 EN 55014 1 EN 55014 2 EN55022 EN 61000 3 2 EN 61000 3 3 EN 61000 ...
Страница 15: ...SS OR ANY OTHER DAMAGES RESULTING FROM THE USE OF THIS PRODUCT Some states do not allow limitation on the length of implied warranties or the exclusion or limitation of incidental or consequential damages This warranty gives you specific legal rights You may have other rights which vary from state to state No individual may accept for or on behalf of Labnet International Inc any other obligation o...