background image

 

11/13/06 

Overview 

The ENDURO 300 Volt Power Supply is a microprocessor-controlled power supply 
designed to meet most electrophoresis needs in a single, easy to use unit.  The power 
supply is capable of running constant voltage / constant current applications 
concurrently. This instrument is ideal for DNA/RNA electrophoresis, SDS-PAGE, native 
PAGE, and second-dimension SDS-PAGE applications.  With four sets of output jacks 
that can be used simultaneously, the ENDURO 300 Volt Power Supply is designed to 
efficiently handle multiple electrophoresis gel tanks and use a small amount of lab space.  
The ENDURO 300 Volt Power Supply offers two modes, Constant Voltage or Constant 
Current Mode.  This manual describes the setup and operation of the ENDURO 300 Volt 
Power Supply including important information on safety and maintenance of the unit. 
 

Description of Buttons and Switches 

 

Front View of ENDURO 300 Volt Power Supply

 

 

Rear View of ENDURO 300 Volt Power Supply 

 

Control Buttons 

LCD Display 

Output 
Jacks 

Power Inlet 

Voltage 
Selector 

Power 
Switch 

Fuse Drawer 

Содержание Enduro E0303

Страница 1: ...11 13 06 300 Volt Electrophoresis Power Supply Catalog Number E0303 110Volt E303 230V 230Volt...

Страница 2: ...rview 6 Description of Buttons and Switches 6 Getting Started 7 Important Guidelines 8 Operation 8 Choosing Limiting Parameter Settings 9 Troubleshooting 11 Repair and Maintenance of 300V Power Supply...

Страница 3: ...the risk of electrical shock to the user Please follow the guidelines below and read this manual in it s entirety to ensure safe operation of the unit The ENDURO 300 Volt Power Supply has been design...

Страница 4: ...upply to indicate an area where a potential shock hazard may exist Package Contents Component Quantity ENDURO 300 Volt Power Supply 1 each Instruction Manual 1 each Extra Fuse 1 each Power Cord 1 each...

Страница 5: ...odes Constant Voltage 1V increment settings Constant Current 1 mA increment settings Crossover Automatic Display type Backlit LCD Graphic type Display size 53 64 x 15 64 mm W x H Pause function Yes Sa...

Страница 6: ...n be used simultaneously the ENDURO 300 Volt Power Supply is designed to efficiently handle multiple electrophoresis gel tanks and use a small amount of lab space The ENDURO 300 Volt Power Supply offe...

Страница 7: ...values Getting Started Installing the ENDURO 300 Volt Power Supply 1 Check the label located near the AC inlet to ensure that the unit is compatible with locally provided voltage 2 Place the ENDURO 30...

Страница 8: ...e Use the START PAUSE and STOP keys to switch on and off the power to the output jacks Recommendation The duration of electrophoresis can be defined in time hours minutes When using this or any electr...

Страница 9: ...most electrophoresis methods resistance increases during the run Limiting the voltage provides the following advantages The same voltage setting can be used regardless of the number or thickness of ge...

Страница 10: ...row keys to set either voltage V or current mA parameters to the appropriate values 4 Press the MODE key to choose the TIME parameter and use the Up Arrow Down Arrow keys to set the duration hours min...

Страница 11: ...is removed from the pre cast gel buffers are prepared correctly and the recommended volume of buffer is added to the electrophoresis unit High voltage application is set to run on a very low current...

Страница 12: ...is supplied with the 300Volt Power Supply For additional fuses contact Labnet s Technical Service department To replace the fuse 1 Turn off the main power switch at the rear of the 300V Power Supply a...

Страница 13: ...run Buffer leaking in a connected system Excessive temperature increase Excessive buffer evaporation Loose connection in a connected system Amperage set to low Change in constant mode Voltage changes...

Страница 14: ...determined by the gel system the resistance can vary over the course of an electrophoretic separation Voltage The velocity with which an ion moves in an electric field will vary in proportion to the...

Страница 15: ...11 13 06...

Страница 16: ...l Inc warrants that this product will be free from defects in material and workmanship for a period of three 3 years from date of purchase If a defect is present Labnet International will at its optio...

Страница 17: ...your responsibility to decontaminate the equipment in case of biological chemical and or radiological contamination so as to protect from health hazards the persons involved in the disposal and recycl...